Trauma to the extremities such wrist, ankle, limb is very common and affects all population groups. It constitutes a significant public health issue. Standard radiography remains the basic imaging tool. However, as a 2-dimensional (2D) imaging modality it lacks sensitivity and specificity. Misdiagnosis rates are known to be high, especially for non-displaced fractures of the scaphoid and talus as well as erosions due to rheumatoid arthritis. Misdiagnosis leads to over treatment and unnecessary loss of productivity and quality of life including 6-12 weeks in a cast. Missed fractures can result in a chronic, non-healing fracture that may require surgical fixation and early arthritis of the joint. From a physician perspective, a missed diagnosis can result in a lawsuit and an expensive settlement/penalty. Computed tomography (CT) offers high resolution and excellent visualization of bone and joint morphology, and Magnetic Resonance Imaging (MRI) delivers soft tissue and cartilage visibility. However, cost, space and workflow related issues make them prohibitive for small orthopedic clinics. Although the radiation dose of a CT scan has been reduced considerably in recent years, it is still significantly higher than a regular radiograph. The whole-body scanners also have difficulties in imaging patients in portable and weight-bearing conditions. Dedicated extremity CT scanners have been commercialized recently in an attempt to address the current deficiency. They still suffer from higher cost and at such have a limited installation base.
The aim of this one-year study is to demonstrate Tomo-E's clinical utility for diagnosis of wrist fractures. Tomo-E is a compact and stationary device that utilizes a distributed carbon nanotube (CNT) x-ray source array that will be specially designed for extremity imaging to collect all the projection views without any mechanical motion.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
30
High-resolution limited-angle tomography positioning and examination will vary depending on injury.
Standard of Care radiographic imaging of wrist.
University of North Carolina at Chapel Hill
Chapel Hill, North Carolina, United States
Sensitivity
The sensitivity of Tomo-E is defined as the ability of readers (radiologists) to detect wrist fractures in patients. Diagnostic accuracy will be defined by the presence of a fracture as clinically determined by the attending orthopedic surgeon. All Tomo-E scans will be reviewed in a standardized reader study at the conclusion of all study imaging to determine the overall sensitivity of the device.
Time frame: 8 months
Specificity
The specificity of Tomo-E is defined as the ability to distinguish between individuals that do not have a wrist fracture. Diagnostic accuracy will be defined by the presence of a fracture as defined clinically by the attending orthopedic surgeon. All Tomo-E scans will be reviewed in a standardized reader study at the conclusion of all study imaging to determine the overall specificity of the device.
Time frame: 8 months