The Imagio® Pivotal Study is intended to evaluate if the results observed in the previous Feasibility Study can be confirmed for pre-specified effectiveness endpoints. Using Intention-to-Diagnose (ITD) masses from the PIONEER Pivotal study, 480 to 840 masses are to be used for the Pivotal Study, the remaining masses will be reserved for training the SenoGram® model.
This will be a controlled, blinded, multi-reader, multi-case (MRMC) study using a sequential design. The study will include 15 readers with an additional 5 as back-up readers depending on qualifications and availability. Readers that participated in the Seno PIONEER Study (NCT01943916) or the Reader-01 Feasibility Study (NCT03708393) are not eligible to participate as readers in this Pivotal Study. Imagio® \[Ultrasound (IUS) + Optoacoustic (OA)\] (IUS)(IUS+OA) training will be completed prior to any reads taking place. Read 1 will be immediately followed by Read 2 within the same read session.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
SINGLE
Enrollment
480
Imagio ultrasound images to be reviewed as part of Reader study
Imagio Ultrasound + Optoacoustic images to be reviewed as part of Reader study
Mammography as available per standard of care
American College of Radiology Center for Research and Innovation
Philadelphia, Pennsylvania, United States
Gain in Specificity at Fixed 98% Sensitivity (fSp)
Primary effectiveness endpoint was the difference (gain) in specificity (fSp) at fixed 98% sensitivity for the Imagio IUS+OA relative to IUS alone, across all 15 independent readers; both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). fSp derived from empirical receiver operating characteristic (ROC) using endpoint interpolation.
Time frame: Baseline to 12 months +/- 30 days follow-up
Negative Likelihood Ratio (NLR)
NLR for the Imagio IUS+OA relative to IUS alone, across all 15 independent readers; both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). NLR = \[(1-sensitivity) / specificity\].
Time frame: Baseline to 12 months +/- 30 days follow-up
Positive Likelihood Ratio (PLR)
PLR for the Imagio IUS+OA relative to IUS alone, across all 15 independent readers; both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). PLR = \[sensitivity / (1-specificity)\].
Time frame: Baseline to 12 months +/- 30 days follow-up
Partial Area Under the Curve (pAUC)
pAUC considers only the region of the area under the ROC curve that corresponds to clinically relevant values of sensitivity, defined as 95% to 100%, for Imagio IUS+OA vs. IUS alone, averaged across 15 readers. Both imaging modalities used in each subject (subject as own control); results for each imaging modality compared to biopsy diagnosis or 12-month follow-up ruling of benign as determined by truth panel (ground truth). A higher pAUC indicates better diagnostic accuracy.