L-PLUS (Lightning PLUS) is the code name of a blood glucose monitoring system (BGMS) consisting of a new meter and marketed CONTOUR PLUS® test strip. This clinical trial will assess the performance (accuracy) of the L-PLUS meter by lay users with diabetes enrolled as subjects in the study, and by health care professionals (also called study staff). The trial will follow the requirements and procedures described in ISO 15197:2013 International Standard.
This study is designed to satisfy the ISO 15197:2013, Section 8 requirements. The study will be conducted at a minimum of one clinical site and enroll a total of approximately 130 persons with diabetes. For this study, there must be at least 120 evaluable results each for fingerstick and AST self-test results, as well as for staff testing subject fingertip and venous blood using the L-PLUS BGMS. Subject enrollment will continue until the required evaluable samples are collected. The subjects will be 18 years and older. Each subject will make one visit, lasting approximately 1 hour at the clinical site. All subjects will complete the informed consent process before performing any study procedures. All Subjects will be assigned a L-PLUS meter and one Contour Plus test strip lots. The untrained subjects will learn to use the L-PLUS meter by reading the instructional materials. Each subject will perform one fingerstick self-test, one palm test, and staff will perform a fingerstick test on the subject. Subjects will be given a venipuncture and the venous blood will be tested by the study staff with the L-PLUS system. All subject capillary and AST self-test results and study staff fingerstick results will be compared to YSI fingerstick results. Venous blood meter tests will be compared to YSI results of the corresponding venous blood. Hematocrit will be measured for all subjects. All Subjects will then complete a questionnaire (Questionnaire 1) to provide feedback on the meter system and instructions for use.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
SUPPORTIVE_CARE
Masking
NONE
Enrollment
130
Blood Glucose Monitoring System
Rainier Clinical Research Center, Inc.
Renton, Washington, United States
Accuracy of BGMS Compared to the Reference Analyzer - Fingerstick Blood Test
Accuracy of BGMS test with fingerstick blood. The results of the fingerstick meter test performed by the participant will be compared to results of the same blood on a laboratory reference analyzer using pre-specified accuracy criteria.
Time frame: Day 1
Accuracy of BGMS Compared to the Reference Analyzer - Palm Alternative Site Test (AST)
Accuracy of BGMS test with palm blood. The results of the palm blood meter test performed by the participant will be compared to results of the same blood on a laboratory reference analyzer using pre-specified accuracy criteria
Time frame: Day 1
Accuracy of BGMS Compared to the Reference Analyzer - Venous Blood
Accuracy of BGMS using venous blood compared to the reference analyzer for all subjects using pre-specified accuracy criteria
Time frame: Day 1