The principal objective of the study is to perform a first step of technical adjustment and preliminary validation of a diagnostic test for hepatitis B (HBsAntigen, anti-HBs Antibody and anti-HBc Antibody) on capillary blood with the device under study in non-immunized subjects against hepatitis B, in subjects with hepatitis B, in subjects with old or cured hepatitis B and in hepatitis B vaccine recipients, compared with the values obtained with the reference technique of medical biology laboratory on serum. The analysis will be performed on the entire cohort.
The diagnosis of hepatitis B in precarious populations is often difficult. The use of a in vitro diagnosis-medical device in point of care could alleviate this difficulty. It would identify whether or not the subjects are infected with the virus, whether they are immunized or not. Then, depending on the results, the subjects could be directly vaccinated or referred to a laboratory of medical biology for confirmation / invalidation of the diagnosis.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
OTHER
Masking
NONE
Enrollment
40
test of capillary blood with the demonstrator at two time and 2 measures and 2 finger by subject.
Clinical research center CIC1406
Grenoble, France
feasibility of correlation
To evaluate the feasibility estimated by correlation of the measurements (HBsAg, anti-HBs Ab, HBc-Ab) between the IVD-MD in point of care on capillary blood study and the laboratory results obtained on serum
Time frame: Day 0
repeatability of the measurement HBsAg, anti-HBs Ab, HBc-Ab
Comparison of the repeatability of the measurement of HBsAg, anti-HBs Ab, HBc-Ab on 2 different fingers with the same device and the same batch of consumables with 2 different times
Time frame: day 0
reproductibility of the measurement HBsAg, anti-HBs Ab, HBc-Ab
Comparison of the reproducibility of HBsAg, anti-HBs Ab, HBc-Ab via a test from 2 drops of blood on 2 different fingers with 2 different cartridges and two different readers
Time frame: day 0
correlation according to status
Comparison of the correlation coefficients of the measurements between the IVD-MD in point of care and the biology automaton according to the different statuses
Time frame: day 0