To establish the safety and effectiveness of the Boston Scientific Cardiac Cryoablation System for treatment of symptomatic, drug refractory, recurrent, paroxysmal atrial fibrillation (AF).
Multi-center, open label, prospective, single arm study to document the safety and performance of Boston Scientific's Cryoablation System. The Cryoablation System is intended for cryoablation and electrical mapping of the pulmonary veins for pulmonary vein isolation (PVI) in the ablation treatment of patients with paroxysmal atrial fibrillation (PAF). All subjects fitting the enrollment criteria, signing the consent and undergoing the index procedure with the study devices will be followed up for twelve (12) months after the index procedure.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
458
The Boston Scientific Cardiac Cryoablation System is intended for ablation and electrical mapping of the pulmonary veins for pulmonary vein isolation (PVI) in the ablation treatment of patients with paroxysmal atrial fibrillation (PAF).The main devices and components of the system consists of the following: * POLARx™ or POLARx™ FIT Cryoablation Catheter * POLARMAP™ Catheter * POLARSHEATH™ * SMARTFREEZE™ Console * Diaphragm Movement Sensor (DMS) * Related Accessories
Percentage of Participants With Freedom From Primary Safety Events at 12 Months Post-procedure (Acute and Chronic Events) Using the Boston Scientific Cardiac Cryoablation System With POLARx Cryoablation Balloon Catheter Models
This measure reports the observed safety event-free rate at 12 Months follow up for all Non Roll-In treatment and intent subjects. The safety events are a composite of procedure-related and device-related adverse events. 7 days (Death, Myocardial infarction, Transient ischemic attack, Stroke/ Cerebrovascular accident, Vascular access complications, Mitral or tricuspid valvular damage, Thromboembolism/ Air embolism leading to a life-threatening event such as a ventricular arrhythmia, stroke, pulmonary embolism, or myocardial infarction and, thromboembolic events that result in permanent injury, require intervention for treatment or prolongs or require hospitalization for more than 48 hours, Gastroparesis/injury to vagus nerve, Pneumothorax, Pulmonary edema/heart failure AV block); 30 days (Cardiac tamponade/perforation); 12 months: (Atrial esophageal fistula, Severe Pulmonary vein stenosis, Persistent phrenic nerve palsy .
Time frame: 12 Months
Percentage of Participants With Freedom From Treatment Failure at 12 Months Post-procedure Using the Boston Scientific Cardiac Cryoablation System With POLARx Cryoablation Balloon Catheter Models
Freedom from the composite of: Failure to achieve acute procedural success in the index procedure or repeat procedure during the blanking period; Use of amiodarone post index procedure; Surgical treatment for AF/AFL/AT post index procedure; Use of non-study ablation catheter for AF targets in the index procedure or repeat procedure during the blanking period; More than one repeat procedure with the POLARx catheter during the blanking period (90 days post index procedure); Documented AF, or new onset of AF or AT event 91-365 days post index procedure; Any of the following for AF, or new onset of AF or AT 91-365 days post index procedure: Repeat procedure - Electrical and/or pharmacological cardioversion for AF/AFL/AT - Prescribed AAD consisting of any Class I/III, including Amiodarone, and any Class II/IV medications taken for control of AF, AT, AFL recurrence
Time frame: 12 Months
Percentage of Participants With Freedom From Primary Safety Events at 3 Months Post Procedure Using the Boston Scientific Cardiac Cryoablation System With POLARx FIT Cryoablation Balloon Catheter Models
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Banner University Medical Center
Phoenix, Arizona, United States
Pima Heart and Vascular
Tucson, Arizona, United States
Arrhythmia Research Group
Jonesboro, Arkansas, United States
Cedars - Sinai Medical Center
Los Angeles, California, United States
Mercy General Hospital
Sacramento, California, United States
Torrance Memorial Medical Center
Torrance, California, United States
Tallahassee Memorial Hospital
Tallahassee, Florida, United States
AdventHealth Tampa
Tampa, Florida, United States
St. Joseph's Hospital
Tampa, Florida, United States
Emory University Hospital
Atlanta, Georgia, United States
...and 34 more locations
Composite of procedure-related and device-related adverse events. 7 days post index procedure or hospital discharge, whichever is later, unless denoted as events counting through 3 months post index procedure (Death, Myocardial infarction, Major Vagal Nerve Injury/Gastroparesis, Transient ischemic attack, Stroke/Cerebrovascular accident, Thromboembolism, Cardiac tamponade/perforation, Pneumothorax, Major vascular access complications, Pulmonary edema/heart failure, AV block, Atrial esophageal fistula, severe pulmonary vein stenosis (≥70% reduction in the diameter of the PV or PV branch from baseline), Persistent phrenic nerve palsy)
Time frame: 3 months
Percentage of Participants With Acute Procedural Success Defined as the Achievement of Electrical Isolation of All PVs by Using the Boston Scientific Cardiac Cryoablation System With POLARx Fit Cryoablation Balloon Catheter Models
The primary effectiveness endpoint is the rate of acute procedural success where acute procedural success is defined as the achievement of electrical isolation of all PVs by using the Cardiac Cryoablation System with the POLARx FIT cryoablation balloon catheter models (with treatment applied at 28 mm or 31 mm balloon size per physician discretion). Electrical isolation of a PV is demonstrated by entrance and exit block.
Time frame: 1 day
Percentage of Participants With Acute Procedural Success Defined as the Achievement of Electrical Isolation of All PVs by Using the Boston Scientific Cardiac Cryoablation System With POLARx Cryoablation Balloon Catheter
The achievement of electrical isolation of all PVs by using the POLARx Cardiac Cryoablation System. Electrical isolation of a PV is demonstrated by entrance and exit block.
Time frame: 1 day
Number of Participants With Freedom From Reportable Adverse Events at 12 Months Post Procedure Using the Boston Scientific Cardiac Cryoablation System With POLARx Cryoablation Balloon Catheter Models
* All Serious Adverse Events * All Study Procedure-Related Adverse Events * All Study Device-Related Adverse Events * All Study Device Deficiencies * Unanticipated Adverse Device Effects/Unanticipated Serious Adverse Device Effects previously not defined in Investigator's Brochure or DFU
Time frame: 12 months
Percentage of Participants With Freedom From Safety Events at 12 Months Post Procedure Using the Boston Scientific Cardiac Cryoablation System Using the Boston Scientific Cardiac Cryoablation System With POLARx FIT Cryoablation Balloon Catheter Models
The following events will be counted through 7 days post index procedure or hospital discharge, whichever is later: Death, Myocardial infarction, Major Vagal Nerve Injury/Gastroparesis, Transient ischemic attack, Stroke/Cerebrovascular accident, Thromboembolism, Pneumothorax, Major vascular access complications, Pulmonary edema/heart failure, AV block. The following event will be counted through 30 days post index procedure: Cardiac tamponade/perforation. The following events will be counted through 12 months post index procedure: Atrial esophageal fistula, Severe pulmonary vein stenosis (≥70% reduction in the diameter of the PV or PV branch from baseline), Persistent phrenic nerve palsy.
Time frame: 12 Months
Percentage of Participants With Freedom From Treatment Failure at 12 Months Post-procedure Using the Boston Scientific Cardiac Cryoablation System With POLARx FIT Cryoablation Balloon Catheter Models
Freedom from the composite of: Failure to achieve acute procedural success in the index procedure or repeat procedure during the blanking period; Use of amiodarone post index procedure; Surgical treatment for AF/AFL/AT post index procedure; Use of non-study ablation catheter for AF targets in the index procedure or repeat procedure during the blanking period; More than one repeat procedure with the Cryoablation System during the blanking period (90 days post index procedure); Documented AF, or new onset of AF or AT event 91-365 days post index procedure; Any of the following for AF, or new onset of AF or AT 91-365 days post index procedure: Repeat procedure - Electrical and/or pharmacological cardioversion for AF/AFL/AT - Prescribed AAD consisting of any Class I/III, including Amiodarone, and any Class II/IV medications taken for control of AF, AT, AFL recurrence
Time frame: 12 Months