The primary objective of the study is to demonstrate the accuracy in healthy adults with Vital USA Device by comparison to a non-invasive (auscultatory) reference sphygmomanometer. The procedure, data collection methods and data analysis that are outlined in the protocol follow the standard: International Standards Organization (ISO) 81060-2:2018 Non-invasive Sphygmomanometers - Part 2: Clinical validation of automated measurement type.
This study will include up to 100 adults ≥ 18 years of age. The final subject database will contain no fewer than 85 subjects with a minimum of 255 paired observations. At least 90% of the subjects will contribute 3 paired observations, if necessary the balance of the data will be added from additional subjects. An interim analysis will be conducted at the end of Part 1 data collection and will consist of simple statistics that will compare the test device tot he reference cuff non-invasive (auscultatory) reference sphygmomanometer. Part 2 data collection will be conducted after Part 1 data collection interim report is reviewed and additional data collection is approved by the client. It will test an additional 55 subjects (plus extra subjects as replacements if necessary).
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
SUPPORTIVE_CARE
Masking
NONE
Enrollment
85
The end goal is to provide Non-invasive Blood Pressure (NIBP) accuracy data to support validation of the Vital Detect blood pressure monitor.
Clinimark Laboratory Services
Louisville, Colorado, United States
Validate Non-invasive Blood Pressure Measurement Accuracy for the Vital Detect Blood Pressure Monitor.
The primary objective of this study is to provide a validation for non-invasive blood pressure measurement accuracy for the Vital Detect blood pressure monitor when compared to the Reference Sphygmomanometer. Data taken was based on dual observers (Criterion 1(C1), Criterion 2 (C2) for both Systolic and Diastolic determinations. Since the same arm sequential method was used, data for a given subject were averaged to create the reference blood pressure determination against the acceptance criteria. The device under test (DUT) passed or failed the determination based on the mean error and standard deviation of the determination. The acceptance criteria: Criterion 1: Mean Error Standard Deviation (Systolic / Diastolic) (Systolic / Diastolic) \</= 5.0 mmHg \</= 8.0 mmHg Criterion 2: Standard Deviation (Systolic) (Diastolic) \</= 5.33 mmHg \</= 6.09 mmHg
Time frame: 1 Hour