Prospective non-randomized parallel-assignment multi-center clinical investigation. The study design includes two subject cohorts: 1) Paroxysmal Atrial Fibrillation Main Study, and 2) Paroxysmal Atrial Fibrillation High Standard Power Substudy. Subjects in the main study cohorts are to be treated using the full range of ablation power settings in the Instructions For Use. Subjects in the High Standard Power Substudy are to be treated in the upper end of the recommended ablation power settings (40-50 Watts).
This clinical investigation is intended to demonstrate the safety and effectiveness of the TactiFlex(TM) Ablation Catheter, Sensor Enabled(TM) (TactiFlex SE) for treating drug-refractory, symptomatic paroxysmal atrial fibrillation. This clinical investigation will be conducted under an investigational device exemption and is intended to support market approval of the TactiFlex SE catheter worldwide. Three hundred and Fifty Five (355) subjects will be enrolled at up to 50 investigational sites worldwide. This clinical investigation is sponsored by Abbott.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
355
Radiofrequency ablation with the TactiFlex SE ablation catheter. The procedure should be performed according to the TactiFlex SE investigational Instructions for Use document using the recommended ablation parameters as noted in the document.
Radiofrequency ablation with the TactiFlex SE ablation catheter. Subjects in the High Standard Power Substudy are to undergo the same study procedures as subjects in the main study, except that ablation power settings of 40-50 Watts are to be used in the left atrium, unless there is a medical reason to use a lower power.
Number of Participants With Device or Procedure-related Serious Adverse Events
Tthe rate of device and/or procedure-related serious adverse events with onset within 7-days of any ablation procedure that uses the TactiFlex SE catheter (initial or repeat procedure performed 31-80 days of initial procedure) that are defined below: * Atrio-esophageal fistula1 * Cardiac tamponade/perforation1 * Death * Heart block * Myocardial infarction * Pericarditis * Phrenic nerve injury resulting in diaphragmatic paralysis * Pulmonary edema * Pulmonary vein stenosis1 * Stroke/cerebrovascular accident * Thromboembolism * Transient ischemic attack * Vagal nerve injury/gastroparesis * Vascular access complications (including major bleeding events) 1. Atrio-esophageal fistula, cardiac tamponade/perforation and pulmonary vein stenosis will be evaluated through 12-months.
Time frame: Through 12 months
KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence
Freedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period will not be considered a treatment failure. One repeat procedure will be allowed during the blanking period. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmic drugs will not count as a therapy failure provided that only previously failed drugs are taken at doses that do not exceed the previously failed dose.
Time frame: Through 12 months
KM Rate of Freedom From Symptomatic Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence
Freedom from documented symptomatic atrial fibrillation/atrial flutter/atrial tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. One repeat procedure will be allowed for ablation 31-80 days after the initial procedure and will not be considered a treatment failure. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmias drugs will not count as a therapy failure provided that only previously failed Class I or III anti-arrhythmias drugs are taken at doses that do not exceed the previously failed dose.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Affinity Cardiovascular Specialists, LLC
Birmingham, Alabama, United States
St. Bernards Medical Center
Jonesboro, Arkansas, United States
Arkansas Heart Hospital
Little Rock, Arkansas, United States
Mills-Peninsula Medical Center
Burlingame, California, United States
Scripps Health
La Jolla, California, United States
University of California at San Diego (UCSD) Medical Center
San Diego, California, United States
South Denver Cardiology Associates PC
Littleton, Colorado, United States
AdventHealth Orlando
Orlando, Florida, United States
Piedmont Athens Regional Medical Center
Athens, Georgia, United States
St. Luke's Regional Medical Center
Boise, Idaho, United States
...and 35 more locations
Time frame: Through 12 months
KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence With Only 1 Ablation Procedure
Freedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. Any repeat ablation procedure in the left atrium will be considered a treatment failure. Failure to achieve acute procedural success during the ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmic drugs will not count as a therapy failure provided that only previously failed Class I or III anti-arrhythmic drugs are taken at doses that do not exceed the previously failed dose.
Time frame: 12 months
KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence, Without Anti-arrhythmic Drugs
Freedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up (9 months after a 90-day blanking period). Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. One repeat procedure will be allowed for ablation 31-80 days after the initial procedure and will not be considered a treatment failure. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, any use of a Class I or III anti-arrhythmic drugs will count as an effectiveness failure.
Time frame: Through 12 months