Hypoglycemia (low blood glucose) is a very common problem in newborns, and has been associated with poor neurodevelopment, cognition, and school performance. At-risk newborns (infants of diabetic mothers \[IDM\], large \[LGA\] and small \[SGA\] for gestational age infants, and late preterm \[LPT\] infants) undergo a hypoglycemia screening protocol that involves numerous intermittent needle sticks to test glucose levels on bedside glucometers; however, continuous glucose monitoring (CGM, currently not approved for clinical use in babies), via a small sensor placed in the thigh (only 1 needle stick), would likely decrease pain while providing continuous glucose levels. This study will evaluate the feasibility, safety, and precision of CGM in at-risk newborns, and determine if this method would decrease the amount of painful procedures and episodes of hypoglycemia missed by intermittent sampling. As part of regular medical care, participants will undergo intermittent blood glucose screening with heel sticks as per the current hospital standard of care protocol. Regular medical care involves checking the participant's blood glucose via heel stick every few hours using a bedside glucometer, with another heel stick to confirm low values in the laboratory. If the participant has low values, he/she may be treated with oral glucose gel, feedings of breast milk or formula, or intravenous (IV) fluids in the Neonatal Intensive Care Unit (NICU). This research study involves placing a CGM device in addition to undergoing the current blood glucose screening protocol and treatment. As soon as possible after birth, a continuous glucose monitoring device (Dexcom G7) will be placed on the participant's thigh by a research team member, and will blindly continuously record glucose levels that will be analyzed after discharge. Everyone who agrees to participate in this study will have placement of this experimental device. The investigational device will stay in place for the same amount of time that a participant is undergoing blood glucose monitoring as per the current standard of care protocol, for a maximum of 7 days. A participant may need to have his/her blood glucose checked after 7 days for regular medical care (and not for research), because his/her glucose concentrations are still low. Being in the research study will not affect a participant's medical care, and will not affect how long he/she needs blood glucose monitoring or treatment. A research team member will place and remove the CGM. Nurses will evaluate the site of the device for signs of irritation, infection, bleeding, and any other issues at least 3 times per day. After discharge from the hospital, data will be collected from the participant's medical record and participant's mother's medical record, including the participant's sex and birth weight, blood glucose values, details of feedings, treatments given for low glucose concentrations, and NICU admission data. Data that will be collected from the participant's mother's medical record includes age and race, prenatal data, medical history, and medication use. The participant's parents will be asked to fill out a short survey about their experience with this device when it is removed.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
100
The Dexcom G7 Glucose Monitoring System (San Diego, CA) reports continuous interstitial blood glucose concentrations every 5 minutes, does not require calibration, and involves only 1 needle stick to place the sensor. The manufacturer recommends that a single sensor may be used for up to 10 days. The Dexcom G7 system is the only FDA approved CGM system for children (age \>2 years).
Penn State Health Milton S Hershey Medical Center
Hershey, Pennsylvania, United States
RECRUITINGAgreement of blood glucose concentrations from CGM vs intermittent monitoring
Bland-Altman analysis to evaluate bias and the agreement interval between CGM and intermittent testing
Time frame: 1-7 days
Likert-scale questionnaire results from parents, nurses, and research team
Feasibility of the CGM device will be evaluated by administering Likert-scale questionnaire results from parents, nurses, and research team. Questionnaires will assess acceptability of using the device by parents, nurses, and researchers; and ease of use, placement, and maintenance of the device.
Time frame: 7 days
Difference in number of needle sticks with CGM vs intermittent glucose monitoring
The total number of estimated needle sticks with GCM will be compared to the total number of needle sticks from intermittent monitoring, using t-test.
Time frame: 1-7 days
Difference in number of episodes of hypoglycemia diagnosed with CGM vs intermittent glucose monitoring
Total number of hypoglycemic episodes (blood glucose concentration \<40 mg/dL at \<4 hours, and \<45 mg/dL thereafter) from GCM (episodes lasting \>10 minutes) will be compared to the total number of hypoglycemic episodes from intermittent monitoring using t-test. Area under curve analysis will be completed to evaluate duration and severity of hypoglycemic episodes from CGM vs intermittent monitoring, as well as the response to treatment (feedings, dextrose gel, IV dextrose boluses, and dextrose infusions) and timing to normalization of glucose concentrations.
Time frame: 1-7 days
Number of participants with adverse outcomes at the device site
Safety of CGM device will be evaluated by examinations for the presence of irritation, infection, cellulitis, bleeding, or other concerns at the device site.