The gekoTM device is a small disposable battery powered device that attaches to the lower leg near the knee and stimulates the common peroneal nerve, causing muscle contraction of the lower leg. This contraction increases blood flow and the device is used to treat several conditions including deep vein thrombosis and venous leg ulcers. There is evidence that some powered muscle stimulators and related devices can affect Cardiac Demand Pacemakers. This study is to ascertain if there is any such interaction between the gekoTM device and permanently implanted cardiac pacemakers.
CFR.890.5850 requires that the labelling for powered muscle stimulators includes the contraindication 'Powered muscle stimulators should not be used on patients with cardiac demand pacemakers'. Further, IEC 60601-2-10:2012 clause 210.7.9.2.101d, requires that the instructions for use provide: 'Advice that a PATIENT with an implanted electronic device (for example a cardiac pacemaker) should not be subjected to stimulation unless specialist medical opinion has first been obtained'. However, no details on what criteria the specialist should apply. There is evidence that some powered muscle stimulators, and related devices can affect Cardiac Demand Pacemakers(1,2). Traditional powered muscle stimulators use pulse trains at a frequency of at least 30Hz(3) to elicit muscle contractions. The electrodes for these devices may be placed at number of sites on the body depending on the application, including the legs, arms and torso, often with well-spaced electrodes that may be attached to different limbs, sending electrical impulses through the torso. The geko™ range of devices, however operate by delivering a single pulse at a frequency of 1Hz with electrodes placed topically, close together, at the knee. The nature of the stimulation pulse, and its location make it extremely unlikely that the geko™ devices could interfere with cardiac demand pacemakers. However, there is no direct evidence to support this assumption. This trial will determine if there are any adverse interactions between the geko™ and cardiac pacemakers.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
PREVENTION
Masking
NONE
Enrollment
28
Assess if cardiac pacemakers detect geko™ electrical output pulse.
University Hospital of Wales
Cardiff, United Kingdom
The Dudley Group NHS Foundation Trust
Dudley, United Kingdom
The primary endpoint is percentage of pacemaker sensed geko™ pulses in a 30 second period.
Interference with pacemaker function.
Time frame: One day - single follow-up.
The secondary endpoints are the incidence of adverse events (AEs), incidence of serious AEs (SAEs), incidence of study treatment related AEs, and the incidence of investigational device related AEs.
Incidence of adverse and serious adverse events.
Time frame: One day-single follow-up.