This is a multicenter, prospective, observational study to evaluate the analytical performance of the Abionic IgE Multi-Allergen Test Panel on the abioSCOPE® device in a U.S. point-of-care environment within a clinical laboratory operating under a CLIA certificate for tests of moderate complexity. The study will assess point-of-care ('external') precision, sample type comparison and correlation with a reference method (Phadia Laboratory System, ThermoFisher Scientific).
Study Type
OBSERVATIONAL
Between-Run Reproducibility K3-EDTA venous whole blood 3 mL 1 venous draw Operator-to-operator imprecision (abioSCOPE) K3-EDTA capillary whole blood 0.05 mL/draw; 0.15 mL total 3 finger sticks (3 different fingers) Sample type comparison (abioSCOPE) K3-EDTA capillary whole blood, venous serum, venous K3-EDTA plasma 0.35 mL (finger stick), 9 mL of whole blood to get a minimum of 3 mL of serum and 9 mL of whole blood K3-EDTA to get a minimum of 3 mL of plasma 1 finger stick; 1 venous draw into serum tube (9 mL) and 1 venous draw into K3-EDTA plasma tube (9 mL), in total 18 mL Method comparison (abioSCOPE and ImmunoCAP/Phadia) K3-EDTA capillary whole blood, venous K3-EDTA plasma 0.05 mL (finger stick), 9 mL of whole blood to get a minimum of 3 ml of plasma 1 finger stick; 1 venous draw into K3-EDTA plasma tube (9 mL tube)
George Washington University, Washington DC
Washington D.C., District of Columbia, United States
Johns Hopkins University, Baltimore
Baltimore, Maryland, United States
The Bernstein Clinical Research Center Cincinnati
Cincinnati, Ohio, United States
POC variance components (in a POL environment):
* External Precision Study ('reproducibility study' 'site-to-site precision study'): Demonstrate that the imprecision of K3-EDTA venous plasma using the IVD CAPSULE Allergic Asthma panel on the abioSCOPE device (all 5 allergens/allergen mixes and total IgE) in the hands of trained Healthcare Professionals in a Physician Office Laboratory (POL) is within the expected range of the imprecision established in the clinical laboratory by Laboratory Scientists. * Operator-to-Operator imprecision on whole blood: Demonstrate that the between-Operator reproducibility of the total IgE of the IVD CAPSULE Allergic Asthma panel on the abioSCOPE device test results is within the expected range of variability. * Between-run reproducibility: Demonstrate that the between-run reproducibility of venous whole blood samples measured with the IVD CAPSULE Allergic Asthma panel on the abioSCOPE device (all 5 allergens/allergen mixes and total IgE) is within the expected range of variability.
Time frame: Day 1
Sample type comparison:
Demonstrate that the IVD CAPSULE Allergic Asthma panel on the abioSCOPE device test results obtained from capillary whole blood correlates well with values obtained from K3-EDTA anticoagulated venous plasma and serum samples.
Time frame: Day 1
Method comparison
Demonstrate that test results obtained with the total IgE test of the IVD CAPSULE Allergic Asthma panel on the abioSCOPE device correlates with total IgE test results from the reference method (Phadia Laboratory System, ThermoFisher Scientific).
Time frame: Day 1