This is a single center, single-arm, quantitative study evaluating the efficacy of the TatchSleep wireless sensor patches as a tool to aid in sleep apnea analysis as compared to an overnight PSG evaluation.
This is a multi-center, single-arm, quantitative study evaluating the effectiveness of the TatchSleep Pro wireless sensors as a tool to aid in sleep apnea diagnosis as compared to an overnight PSG evaluation. Investigators will identify approximately 50 eligible patients who have been recommended to receive an overnight PSG for the detection/evaluation of sleep apnea. After informed consent is obtained, a brief sleep-related medical history will be collected including relevant demographics. Females of childbearing potential will be asked to undergo a urine dipstick pregnancy test to determine their eligibility for inclusion in the study. Patients will undergo their PSG test while simultaneously wearing 2 TatchSleep Pro patches and an FDA-cleared pulse oximeter compatible with and connected to the TatchSleep Pro system for a single night at the clinic. The same make and model of pulse oximeter will be used for all subjects. Patients will be observed overnight by trained sleep technicians who will setup the TatchSleep Pro device and collect the sleep data via the companion smartphone application. A follow-up communication will be made with patients within 5 days after the sleep study to assess any adverse events. Following the data collection, sleep data from PSG and TatchSleep Pro will be scored by an independent qualified sleep technologist (the primary reader) to yield an analysis of the accuracy of TatchSleep Pro compared with the PSG signals. At least one and up to two additional readers (secondary readers) will score only the TatchSleep Pro data to obtain an estimate of inter-rater reliability.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
53
Wearable measurement device
Montefiore Medical Center
The Bronx, New York, United States
Correlation Between Polysomnography (PSG) and Tatch Apnea/Hypopnea Index (AHI)
The correlation between AHI (3% desaturation criterion) measured by the TatchSleep Pro device and in-lab PSG will be assessed. Agreement between the two modalities will be analyzed using Bland-Altman analysis, and Pearson correlation coefficients will be calculated for AHI and ODI (3%). This outcome evaluates the concordance between TatchSleep Pro and PSG for key respiratory indices.
Time frame: 12 hours
Percent Agreement of AHI Severity Classification Between TatchSleep Pro and Polysomnography
Agreement in AHI severity classification (Normal, Mild, Moderate, Severe) between TatchSleep Pro and in-lab PSG was assessed using blinded scoring. Severity assignments from both devices were compared using percent agreement.
Time frame: 12 hours
Agreement Between TatchSleep Pro and PSG on AHI Severity Using Cohen's Kappa
Agreement in AHI severity classification (Normal, Mild, Moderate, Severe) between TatchSleep Pro and in-lab PSG was assessed using blinded scorers. The level of agreement was quantified using the weighted Cohen's kappa coefficient.
Time frame: 12 hours