The purpose of this study is to evaluate the safety and collect data on the initial effectiveness of a 1.1mm flexible single-use cryoprobe with oversheath used for transbronchial lung biopsy via a bronchoscopic approach.
A smaller (1.1mm), flexible, single-use cryoprobe with an oversheath has been developed that can be used for transbronchial biopsies. This device has the potential to gather larger and higher quality tissue samples than the standard method using forceps, and with potentially fewer complications than older, larger versions of the cryoprobe. This study does not involve randomization or assigning different patients to different procedures to compare. Everyone who enrolls in this study will undergo a standard of care bronchoscopy with a transbronchial biopsy using the 1.1mm sheath cryoprobe. The only difference from the standard of care approach is that the physician performing the transbronchial biopsy will use the 1.1mm sheath cryoprobe rather than forceps or another currently available larger cryoprobe. Patients enrolled in this study will have data collected by research staff for up to 30 days after the bronchoscopic biopsy procedure is performed. This 30-day follow-up period is the standard of care following bronchoscopic biopsy procedures.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
51
ERBE 1.1mm flexible single-use cryoprobe with oversheath
Johns Hopkins Hospital
Baltimore, Maryland, United States
Vanderbilt University Medical Center
Nashville, Tennessee, United States
Number of Participants With Serious Adverse Events
The number of participants enrolled who experienced a device related Serious Adverse Events (SAEs). SAEs include Grade 3-4 bleeding (bleeding causing cardiopulmonary instability or requiring inflation of a bronchial blocker), pneumothorax requiring chest tube placement (Grade 2+), or 30-day respiratory failure and death.
Time frame: Within 30 days of procedure
Number of Participants With Minor Adverse Events
The number of participants enrolled who experienced a device related Minor Adverse Event (AE). AEs include Grade 1-2 bleeding (bleeding requiring suction to clear or wedging of the biopsied segment with the flexible bronchoscope and/or iced saline), or pneumothorax not requiring chest tube placement (Grade 1).
Time frame: Within 30 days of procedure
Mean Histologic Accessibility Grade
Grade 0: Does not contain alveolar structures and can therefore not be assessed Grade 1: Very poor specimen quality; not possible to assess the relevant morphologic and histologic features Grade 2: Poor specimen quality; assessment of relevant morphologic and histologic structures and features is severely compromised and not possible Grade 3: High limitations in specimen quality; assessment is severely compromised but limited evaluation is possible Grade 4: Moderate limitations in specimen quality; assessment is moderately compromised but evaluation is possible Grade 5: Low limitations in specimen quality; assessment is somewhat compromised but possible Grade 6: The specimen allows for complete and unrestricted assessment of all relevant morphologic and histologic structures and features
Time frame: Data assessed at the time of review of biopsy sample (up to one year post index procedure) and participants were assessed for the outcome measure at the time of biopsy procedure (up to 1 hour)
Diagnostic Yield as Assessed by Number of Patients for Which Cryobiopsy Led to a Diagnosis
Diagnostic yield is defined as number of patients for which cryobiopsy led to a diagnosis. Diagnostic yield will be determined from the results of cryobiopsy specimen only. A biopsy that results in a specific diagnosis, either malignant or benign (granuloma, inflammation, fibrosis, infection) will be assumed to be a true positive. Atypia, minimal inflammation or lung parenchyma without pathologic findings on final pathology reads are considered non-diagnostic.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: During procedure, up to 1 hour
Total Histologic Area (Square Millimeters)
Amount of total histologic tissue in square millimeters observed under microscope.
Time frame: Data assessed at the time of review of biopsy sample (up to one year post index procedure) and participants were assessed for the outcome measure at the time of biopsy procedure (up to 1 hour)
Alveolated Area (Square Millimeters)
Total amount of area in square millimeters containing alveoli.
Time frame: Data assessed at the time of review of biopsy sample (up to one year post index procedure) and participants were assessed for the outcome measure at the time of biopsy procedure (up to 1 hour)
Open Alveoli Percent
Percentage of all alveoli that are open.
Time frame: Data assessed at the time of review of biopsy sample (up to one year post index procedure) and participants were assessed for the outcome measure at the time of biopsy procedure (up to 1 hour)
Percent Crush Artifact
Percent of tissue distortion resulting from even the minimal compression of the tissue, which may rearrange tissue morphology.
Time frame: Data assessed at the time of review of biopsy sample (up to one year post index procedure) and participants were assessed for the outcome measure at the time of biopsy procedure (up to 1 hour)
Artifact Free Lung Parenchyma Percent
Percent of the lung parenchyma (alveoli, alveolar ducts and respiratory bronchioles) without artifact.
Time frame: Data assessed at the time of review of biopsy sample (up to one year post index procedure) and participants were assessed for the outcome measure at the time of biopsy procedure (up to 1 hour)
Activation Time (Seconds)
Time of activation of cryoprobe.
Time frame: At the time of procedure, up to 1 hour
Procedure Time
Time for entire procedure measured in minutes.
Time frame: At the time of procedure, up to 1 hour