Brief Summary: Post-operative morbidity remains a reality as shown by the International Surgical Outcomes Study published in 2016 and 2019 and by several recent publications which focus mainly on hypotension and cardio-vascular complications. Other complications, such as respiratory depression, are less often studied. The hypothesis is that this connected patch could be used in surgical departments to detect a postoperative complication. Biobeat Technologies Ltd has developed a sensor which continuously records the photoplethysmographic waveform and allows the calculation of several physiological parameters: heart rate (HR), oxygen saturation (SpO2), systolic blood pressure (SBP), diastolic blood pressure (DBP), stroke volume (SV), cardiac output (CO), cardiac index (CI), systemic vascular resistance (SVR), respiratory rate (RR), and temperature. The objective of this study is the quantification of hemodynamic, respiratory and temperature abnormalities detected by routine monitoring (routine nursing follow-up) and continuous monitoring by the Biobeat patch during the first 72 postoperative hours.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
109
The Biobeat device (BB-613WP) is a class IIa medical device and is a non-invasive patch used for monitoring vital signs in clinical and non-clinical settings. The sensor continuously records the photoplethysmographic waveform that allows the calculation of several physiological parameters: heart rate (HR), SPO2, systolic blood pressure (SBP) and change in SBP, diastolic blood pressure (DBP) and change in DBP, stroke volume (SV), cardiac output (CO), cardiac index (CI), systemic vascular resistance (SVR), and respiration rate (RR). Translated with www.DeepL.com/Translator (free version)
Groupe hospitalier Paris Saint-Joseph
Paris, France
Hopital Foch
Suresnes, France
Proportion of patients with a hemodynamic abnormality using the data obtained using the Biobeat patch.
A hemodynamic abnormality is defined as an mean Blood Pressure (MBP) \< 60 mmHg
Time frame: 72 hours
Proportion of patients with hypotension detected by the routine nursing follow-up.
A hemodynamic abnormality is defined as: * An MBP \< 70 mmHg * or an MBP \< 65 mmHg * or an MBP \< 80% of the value measured during the preanesthetic consultation * or an MBP \< 70% of the value measured during the preanesthetic consultation * or an MBP \< 80% of the value measured in the OR before the induction of anesthesia * or an MBP \< 70% of the MBP measured in the OR before the induction of anesthesia * or an MBP \> 100 mmHg * or an MBP \> 110 mmHg * or an MBP \> 120 mmHg * or an MBP \> 130 mmHg * or an MBP \> 120% of the MBP measured during the preanesthetic consultation * or an MBP \> 130% of the MBP measured during the preanesthetic consultation * or an MBP \> 120% of the MBP measured in the OR before induction of anesthesia * or an MBP \> 130% of the MBP measured in the OR before induction of anesthesia * or a heart rate \< 40/min * or a heart rate \> 100/min * or a rhythm disorder
Time frame: 72 hours
Proportion of patients with a postoperative respiratory abnormality using the data obtained using the Biobeat patch.
A postoperative respiratory abnormality is defined as: * a breathing rate \< 8/min * or a breathing rate \> 20/min * or arterial oxygen saturation \< 95% * or arterial oxygen saturation ≤ 92% * or arterial oxygen saturation ≤ 90% * or arterial oxygen saturation ≤ 85%
Time frame: 72 hours
Proportion of patients with a postoperative respiratory abnormality detected by the routine nursing follow-up.
A postoperative respiratory abnormality is defined as: * a breathing rate \< 8/min * or a breathing rate \> 20/min * or arterial oxygen saturation \< 95% * or arterial oxygen saturation ≤ 92% * or arterial oxygen saturation ≤ 90% * or arterial oxygen saturation ≤ 85%
Time frame: 72 hours
Proportion of patients with a temperature abnormality using the data obtained using the Biobeat patch.
A temperature abnormality is defined as: * a temperature ≤ 36.8 °C * or a temperature ≥ 38°C * or a temperature ≥ 39°C
Time frame: 72 hours
Quantify and compare the frequency of temperature abnormality detected by the routine nursing follow-up.
temperature abnormality is defined as: * a temperature ≤ 36.8 °C * or a temperature ≥ 38°C * or a temperature ≥ 39°C
Time frame: 72 hours
Concerning the Biobeat patch, frequency of artifacts or of absence of data
An artefact is defined as * a value that is \> 50% different from the previous value, unless it is followed by a value equal to ± 25%; * or a value that is outside the physiologically plausible range (heart rate \< 5 or \> 250 bpm, systolic blood pressure \< 20 mm Hg or \> 300 mm Hg or lower than diastolic pressure plus 5 mm Hg, diastolic blood pressure \< 5 mm Hg or \> 225 mm Hg, change in SpO2 ≥ 8% between two consecutive measurements, breathing rate \< 3 or \> 125 breaths per minute, change in skin temperature ≥ 1° between two consecutive measurements).
Time frame: 72 hours
Collection of postoperative complications that occurred during the monitoring period
A postoperative complication will be classified according to the classification of Dindo and Clavien (Dindo, Demartines et al. 2004).
Time frame: 72 hours
Evaluation of patients' tolerance to wearing the Biobeat device
4- point Likert scale (from 0 = intolerable to 4 = no problem at all)
Time frame: 72 hours