This study compares the sterility of rigid nasoendoscopies after being sterilized with a conventional autoclave machine and a novel rig-S device that uses standard high level disinfectant
There is a standard operating procedure in rigid endoscopy cleaning, and it is performed after each clinical examination. The conventional autoclave is no longer used in the setting of an otorhinolaryngology (ORL) clinic as it is cumbersome and takes a lot of time, and also only cleans one endoscope at a time. There is no standardisation or device that has been reported, and cross contamination has been reported previously. The overturn of patients in the ORL clinic is high, and most practitioners chooses high level disinfectant as a method of sterilising the endoscopes. However, there is no proper device to place these disinfectants in, and results in many cases of broken scopes and cross contamination. This study compares the sterility of the rigid endoscopes after being sterilised by the high level disinfectant that is placed in a novel rig-S device with the gold standard conventional autoclave. The investigators selected all the rigid endoscopes performed in our ORL clinic and randomised them before subjecting them for a swab test for bacterial and fungal, and also used a test kit to test the presence of Hepatitis B.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
OTHER
Masking
SINGLE
Enrollment
500
A new device created to hold the high level disinfectant and automatically indicate when sterilisation process has been completed
The gold standard device to clean all medical instruments
Universiti Kebangsaan Malaysia
Cheras, Kuala Lumpur, Malaysia
Microbial and Hepatitis B test
Rigid endoscopes are swabbed and tested using a hep-B kit
Time frame: 3 days
Acidity level
Acidity level of the high level disinfectant is tested for every 10 cycles of scope
Time frame: 4 hours