A device has been developed which is placed under the mattress, thereby escaping contact with patient, which enables monitoring of sleep activity and has been shown in some studies to provide an estimate of the frequency of sleep disordered breathing. If the estimates of sleep disordered breathing from this new device are medically equivalent to current excepted home sleep apnea test devices within clinically important ranges of sleep disordered breathing, it may have several benefits. This study is designed to help answer several important clinical questions, namely, is the new device clinically equivalent to an established HSAT device for patients who have at least moderately severe OSA, and what is the effect of multiple nights on classification of severity and presence of OSA.
The design will be a prospective test of two diagnostic testing strategies. The tests will take place in the same patient concurrently, so the patient's results from one test will serve as the comparator with the other test device. No randomization or blinding is needed. Patients will be recruited from the Mayo Clinic Center for Sleep Medicine (and another sleep center with enriched racial variety). Consecutive patients in whom the clinician has ordered a WatchPAT™ will be potential candidates for an opportunity to participate in the study. Patients will be made aware of the trial by invitation from members of the healthcare team. Interested patients will be evaluated according to the inclusion and exclusion criteria.
Study Type
OBSERVATIONAL
Enrollment
60
Under-mattress sleep monitor for patients with a confirmed diagnosis of Obstructive Sleep Apnea.
Mayo Clinic in Rochester
Rochester, Minnesota, United States
The apnea hypopnea index (AHI)
The apnea hypopnea index (AHI) provided by one-night of testing with the WatchPAT™ is equivalent to the AHI provided by the mean of 7 nights use of the Withings™ sleep monitor. The equivalence margin will be an AHI≤5/hr.
Time frame: 12 months
Sensitivity
The Withings sleep monitor will have a sensitivity ≥ 0.825 to identify patients with moderate to severe OSA as assessed by a WatchPAT AHI≥15.
Time frame: 12 months