This PMCF study is designed as prospective, multi-center study to collect real-life data. The rationale of this study is to confirm and support the clinical safety and performance of the aforementioned products in 200 patients who will undergo an endovascular intervention within standard-of-care (SOC) where at least 1 of the following devices from AndraTec were used: The LOKUM-Quest guidewire, The LOKUM AMPLATZ guidewire, the Slider Hydrophilic Nitinol Guidewire, the AltoSa-XL PTA balloon, the AltoSa-XL GEMINI PTA Balloon, the OPTIMUS CoCr Stent , the OPTIMUS-CVS PTFE-Covered stent.
Study Type
OBSERVATIONAL
1 or more AndraTec devices (Lokum L-Quest guidewire, Lokum Amplatz guidewire, Slider Hydrophilic Nitinol Guidewire, AltoSa-XL PTA Balloon, Alto-Sa-XL Gemini Balloon, Optimus-CoCr stent, Optimus-CVS PTFE covered stent)
Freedom from MACE (Major Adverse Clinical Events) and/or Device related events during the procedure
MACE defined as: Death, Myocardial Infarction, Stroke, Emergent surgical revascularization of the target vessel, repeat vascularization of the target vessel, bleeding complication requiring transfusion)
Time frame: during procedure
Technical success rate
defined as successfully introduction and deployment of the Lokum L-Quest, and/or the Lokum Amplatz, and/or the Slider Hydrophilic Nitinol Guidewire, and/or the Optimus CoCr Stent, and/or the Optimus PFTE covered BE Stent, and/or the AltoSa-XL-PTA, and/or the AltoSa-XL-Gemini Balloon Catheter according to the respective IFU and without device related deficiencies.
Time frame: during procedure
For the Optimus CoCr and Optimus-CVS PTFE-covered stent: Freedom from MACE post-procedure, after 6- and 12-months.
MACE defined as: Death, Myocardial Infarction, Stroke, Emergent surgical revascularization of the target vessel, repeat vascularization of the target vessel, bleeding complication requiring transfusion)
Time frame: Through study completion, an average of 1 year
For the Optimus CoCr and Optimus-CVS PTFE-covered stent: Freedom from Device-related serious adverse events (SAE's) post-procedure, after 6- and 12-months
Freedom from Device-Related serious adverse events (SAE's)
Time frame: Through study completion, an average of 1 year
Freedom from Target Lesion Revascularization post-procedure, at 6- and 12-months
Defined as freedom from a repeat intervention to maintain or re-establish patency within the region of the treated arterial vessel plus 5mm proximal and distal to the treated lesion edge.