The purpose of the study is to evaluate the diagnostic performance of a Raman spectroscopy analysis device - Spectra IMDx for the detection of gastric cancer and precancerous lesions in vivo.
This is a prospective, single centre study to assess the diagnostic performance of Spectra IMDx system in detection of gastric cancer and precancerous lesions in subjects with suspected gastric intraepithelial neoplasia who are undergoing upper gastrointestinal endoscopic procedures. The Spectra IMDx system comprises a laser system, a spectrometer, a computer with an analysis algorithm installed, and other ancillary parts. The Spectra IMDx system also comprises a Raman probe that can be connected with the main system. The Raman probe is an assembly of optical fibres and optical components arranged for maximal transmission of light energy. When in use, the laser system will emit a 785nm near infra-red laser that will be transmitted through the Raman probe to the distal end. When the laser is interrogated upon a tissue surface, the light energy is absorbed and reflected. The reflected energy is then collected from the distal end of the Raman probe, transmitted back to the main system, and passed through the spectrometer. The collected signal is then processed to obtain the clean Raman signal, which is then parsed through the analysis algorithm to evaluate its risk for cancer (high/low). Feedback will then be generated on the GUI in real time.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
185
The Spectra IMDx system is intended to be used in conjunction with conventional white light endoscopy in upper gastrointestinal endoscopic procedure to identify patients with high risk of developing and/or having gastric cancer and it is intended for use on patients above and equal to the age of 18.
Zhongshan Hospital Xiamen University
Xiamen, Fujian, China
RECRUITINGSensitivity of Spectra IMDx system
To assess the sensitivity of Spectra IMDx system for the detection of gastric cancer and precancerous lesions in vivo in the presence of using histological diagnosis as the gold standard. Diagnostic measurements will be assessed in this trial. Using the histopathological diagnosis as the gold standard, true positives (TPs), true negatives (TNs), false positives (FPs) and false negatives (FNs) can be obtained for both Spectra IMDx (IMDx) system. Sensitivity = TP/(TP + FN).
Time frame: During the procedure
Specificity of Spectra IMDx system
To assess the specificity and accuracy of Spectra IMDx system for the detection of gastric cancer and precancerous lesions in vivo in the presence of using histological diagnosis as the gold standard. Specificity = TN/(TN + FP).
Time frame: During the procedure
Accuracy of Spectra IMDx system
To assess the accuracy of Spectra IMDx system for the detection of gastric cancer and precancerous lesions in vivo in the presence of using histological diagnosis as the gold standard. Accuracy = (TP + TN)/(TP + TN + FP + FN).
Time frame: During the procedure