The objective of this study is to calibrate the blood pressure measurements of the Pulsewave Health Monitor (PAD-2A) device to be within 5±8 mmHg of the average of the dual-observer measurements via an aneroid sphygmomanometer (reference device), as per the methodology described in the ISO 81060-2: AMD\_2020 protocol.
The plan for this study is to calibrate the blood pressure measurements of the PAD-2A device with the dual-observer blood pressure measurements via an aneroid sphygmomanometer (reference device), as per the ISO 81060-2: AMD\_2020 protocol. The data will be collected and analyzed in order to calibrate the PAD-2A device blood pressure measurements to be within 5±8 mmHg of the average of the dual-observer auscultatory blood pressure measurements by adjusting the device algorithm.
Study Type
OBSERVATIONAL
Enrollment
200
This is an observational, comparison study of blood pressure measurements. The objective of this study is to calibrate the blood pressure measurements of the Pulsewave Health Monitor (PAD-2A) device to be within 5±8 mmHg of the average of the dual-observer auscultatory blood pressure measurements via an aneroid sphygmomanometer (reference device), as per the ISO 81060-2: AMD\_2020 universal protocol.
Horizon Health Network
Saint John, Canada
RECRUITINGSystolic Pressure
Non-invasive Pulsewave Health Monitor (PAD-2A) wrist cuff blood pressure device (mmHg)
Time frame: From the start of cuff inflation to the end of cuff deflation (approximately 60 seconds)
Systolic Pressure
Simultaneous auscultatory blood pressure measurements by 2 trained observers
Time frame: From the start of manual inflation of the upper arm blood pressure cuff to the end of (manual) deflation of the upper arm blood pressure cuff (approximately 30-60 seconds)
Diastolic Pressure
Non-invasive Pulsewave Health Monitor (PAD-2A) wrist cuff blood pressure device (mmHg)
Time frame: From the start of cuff inflation to the end of cuff deflation (approximately 60 seconds)
Diastolic Pressure
Simultaneous auscultatory blood pressure measurements by 2 trained observers
Time frame: From the start of manual inflation of the upper arm blood pressure cuff to the end of (manual) deflation of the upper arm blood pressure cuff (approximately 30-60 seconds)