Multi-center trial to assess the feasibility and safety of the EchoMark LP and the EchoMark diagnostic ultrasound system for assessing AV fistula blood flow, diameter, and depth.
This is a prospective, multi-center, single-arm, non-blinded clinical trial designed to evaluate the feasibility and safety of the EchoMark and the EchoSure in subjects undergoing new upper arm autologous arteriovenous fistula creation who require hemodialysis. All subjects will provide informed consent prior to undergoing any study procedures. The study will consists of multiple follow-up visits during the 52 week duration. Subjects are able to re-consent and remain in follow-up for an additional year.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
20
The EchoMark is to be secured a minimum of 3 cm away from the anastomosis. The follow-up EchoSure scans will be conducted by the designated EchoSure user without the investigator and without the sonographer/RVT conducting the Duplex exam in the room, either during the routine work-up prior to the surgeon seeing the patient or after. The EchoSure results for flow, and vessel depth and diameter are shown as numerical values on the screen of the EchoSure device.
Trinity Research Group
Dothan, Alabama, United States
SKI Vascular Center
Peoria, Arizona, United States
Michgan Vascular Center
Flint, Michigan, United States
Surgical Specialist of Charlotte
Charlotte, North Carolina, United States
Primary Safety Endpoint
Composite of new major device related adverse events (MAEs) at 30 days post EchoMark implantation, as adjudicated by the Clinical Events Committee (CEC) including: * Device or procedure-related death * Device related infection * Device related interventions * Device related hospitalizations * Fistula failure * Fistula rupture * Aneurysm
Time frame: 30 Days
Primary Feasibility Endpoint
Rate of Technical Success defined as the successful implantation of the EchoMark implant. Technical success will be assessed from baseline to 4 months.
Time frame: 4 Months
Rate of Technical Success Defined as the Successful Ability to Determine Blood Flow, Diameter, and Depth Measurements Using the EchoSure Diagnostic Ultrasound System. Technical Success Will be Assessed From Baseline to 4 Months.
Rate of Technical Success defined as the successful ability to determine blood flow, diameter, and depth measurements using the EchoSure diagnostic ultrasound system. Technical success will be assessed from baseline to 4 months.
Time frame: 4 Months