Whole-body plethysmography (WBP) is the gold standard for measuring lung volume, but its clinical application is limited because it requires expensive equipment and is complicated to use. Studies have shown that the single-breath helium dilution (SBHD) method, which is commonly used in clinical practice, significantly underestimates lung volume in patients with obstructive lung disease (OLD). Therefore, by comparing the differences in lung volume measured by the SBHD method and WBP in patients with different severities of obstructive lung disease, we aim to establish a correction equation for the SBHD method to determine the total lung volume in obstructive lung disease patients.
Study Type
OBSERVATIONAL
Enrollment
628
The patients with obstructive lung disease simultaneously subjected to the single-breath helium dilution method and Whole-body plethysmography tests.
Key Laboratory of Respiratory Disease of Zhejiang Province, Department of Respiratory and Critical Care Medicine, Second Affiliated Hospital of Zhejiang University School of Medicine
Hangzhou, Zhejiang, China
Establishment of the TLC-SBHD correction equation
TLC-adjusted SBHD (L) = -0.669 + 0.756 × TLC-SBHD (L) - 0.047 × FEV1/FVC + 0.039 × height (cm) - 0.009 × weight (kg) (r2 = 0.753 and adjusted r2 = 0.751)
Time frame: 2 year