Objective: To assess the efficacy and safety of an insulin-glucagon dual-hormone (DH) closed-loop system compared with an insulin-only single-hormone (SH) closed-loop system in adolescent with type 1 diabetes. Methods: In a 26-h, randomized, crossover, inpatient study, 20 children and adolescents with type 1 diabetes used two modes of the DiaCon Artificial Pancreas system: DH and SH closed-loop control. During each study period, participants will have one overnight stay, received three meals and performed exercise for 45 min (bicycle with estimated 50% V02max). Endpoint: The primary endpoint is sensor-derived percentage of time in hypoglycemia (\<3.9 mmol/L).
Upon arrival at the research facility at Steno Diabetes Center Copenhagen, the participant's own pump and CGM are disconnected, and the two study pumps are attached. For dual-hormone control, the study pumps will be filled with FiAsp® and GlucaGen®. For single-hormone control, the study pumps have been filled with FiAsp® and isotonic saline. In addition, participants are provided with a blinded wearable activity and sleep monitoring device (ActiGraph GT9X Link, Pensacola, FL) for activity level estimation and sleep assessment during study participation. The Actigraph is connected to a Bluetooth® Polar heart rate monitors during the exercise session. Female participants deliver a urine sample for pregnancy testing. A sampling cannula is placed in an antecubital vein. The CGM will be calibrated with fingerpick glucose meter (Contour next®, Ascensia Diabetes Care) before initiating the closed-loop control. Even though it is not needed to calibrate the sensor, the accuracy of the sensor is better after one calibration. At day 1 17:00 the study is initiated, and the closed-loop system takes over glucose control. Except from the control approach (single- vs. dual-hormone), the study days are identical. During the study visit, participants can move around freely, but they can only perform actual exercise during the stationary bike exercise session at day 2 16:30. At nighttime, the participants are encouraged to be in bed and, if possible, sleep. At predefined timepoints, the investigators measure blood pressure and pulse, and the investigators ask the participants to rate their nausea level on a visual analog scale (VAS) from 1 to 100 to assess possible side effects of glucagon (vomiting rates as 100). In addition, the investigators ask them to self-rate blood glucose levels before bedtime (23:00), upon wake-up (07:00-08:00), before each meal, before exercise, and every 10 minutes during exercise. Single-hormone and dual-hormone closed-loop control study sessions end 26 hours after study start. The study pumps and the CGMs are disconnected and the participants reconnect their own pumps and CGMs. The two study sessions are separated by at least 36 hours.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
11
Glucagon is filled in the pump and given automatically in case of hypoglycemia or pending hypoglycemia.
Closed-loop system comprise of two DANA RS (R) insulin pumps (FiAsp-GlucaGen vs FiAsp-saline), one DexCom G6 sensor, and a smartphone for the control algorithm.
Steno Diabetes Center Copenhagen
Gentofte Municipality, Denmark
Herlev Hospital
Herlev, Denmark
Percentage of time with glucose values < 3.9 mmol/l as measured by the continuous glucose monitor
Percentage
Time frame: 26 hours during closed-loop control
Number of carbohydrate interventions to treat hypoglycemia
Time frame: 26 hours during closed-loop control
Percentage of time with glucose values in the range 3.9-10.0 mmol/l measured by continuous glucose monitor and plasma glucose
Percentage
Time frame: 26 hours during closed-loop control
Percentage of time with glucose values < 3.9 mmol/l as measured by plasma glucose
Percentage
Time frame: 26 hours during closed-loop control
Percentage of time with glucose values in the range > 13.9 mmol/l measured by continuous glucose monitor and plasma glucose
Percentage
Time frame: 26 hours during closed-loop control
Percentage of time with glucose values < 3.0 mmol/l as measured by continuous glucose monitor and plasma glucose
Percentage
Time frame: 26 hours during closed-loop control
Mean blood glucose value measured by continuous glucose monitor and plasma glucose
mmol/L
Time frame: 26 hours during closed-loop control
Number of hypoglycemic episodes < 3.9 mmol/l on continuous glucose monitor and plasma glucose
No.
Time frame: 26 hours during closed-loop control
Continuous glucose monitored glycemic variability measured as SD
mmol/L
Time frame: 26 hours during closed-loop control
Continuous glucose monitored glycemic variability measured as CV
percentage
Time frame: 26 hours during closed-loop control
Composite outcome: Percentage of participants achieving (1) time in range (3.9-10) > 70 %, (2) time in alert hypoglycemia (<3.9 mmol/l) < 4 %, and (3) time in clinical hypoglycemia (<3.0 mmol) < 1% as measured by CGM and YSI
percentage
Time frame: 26 hours during closed-loop control
Total insulin dose
units
Time frame: 26 hours during closed-loop control
Total glucagon dose
mg
Time frame: 26 hours during closed-loop control
Number of manual insulin boluses
No.
Time frame: 26 hours during closed-loop control
Number of adverse events - Nausea
No of event if visual analog scale (0-100) increase \>10 from baseline
Time frame: 26 hours during closed-loop control
Number of adverse events - Headache
No of event if visual analog scale (0-100) increase \>10 from baseline
Time frame: 26 hours during closed-loop control
Number of adverse events - Palpitation
No of event if visual analog scale (0-100) increase \>10 from baseline
Time frame: 26 hours during closed-loop control
Number of vomits
No of event if visual analog scale (0-100) increase \>10 from baseline
Time frame: 26 hours during closed-loop control
Difference between actual and participant-estimated carbohydrate content in meals
g per meal
Time frame: 26 hours of closed-loop glucose control
Mean Borg scale
Scale of perceived exertion from 6 (no effort activity) to 20 (max effort activity)
Time frame: During 45 minutes exercise
Physical activity intensity measured by ActiGraph GT9X Link
Percentage of sedentary activity
Time frame: 26 hours
Sleep efficiency measured by ActiGraph GT9X Link
The ratio of total sleep time to time in bed
Time frame: 26 hours