The application of conventional endoscopy in remote and outdoor areas lacking facilities remains challenges. Thus, the investigators developed a novel portable upper gastrointestinal endoscopy system that has the same functions as conventional endoscopy. A total of 24 participants from a medical unit on a remote island in China underwent endoscopy with the portable system between March and June 2021. The portable system packed into a suitcase is 68 × 42 × 32 cm in size, weighing less than 35 kg, and comprises a disposable sheathed system.
The application of conventional endoscopy in remote and outdoor areas lacking facilities remains challenges. Thus, the investigators developed a novel portable upper gastrointestinal endoscopy system that has the same functions as conventional endoscopy. A total of 24 participants from a medical unit on a remote island in China underwent endoscopy with the portable system between March and June 2021. The portable system packed into a suitcase is 68 × 42 × 32 cm in size, weighing less than 35 kg, and comprises a disposable sheathed system. The examination time and any malfunctions in the portable endoscopy system during the examination were recorded. After the examination was finished, the disposable sheathed system was removed, and a pressure leak test was performed to check the seal of the system. The patients were followed up within three days to confirm any adverse events such as nausea, vomiting, throat discomfort, bleeding, etc.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
24
Patients arrived at the hospital in the morning after an overnight fast (\>6 hours). The patients took oral dyclonine glue 15 mins before their examination. The doctors prepared the portable endoscopy system so that it was in working status and the disposable sheathed system was placed on the endoscope optical probe.
Changhai Hospital
Shanghai, Shanghai Municipality, China
The examination time
duration from the time the endoscope was inserted into the patient's mouth to the time it was withdrawn.
Time frame: 24 hours
adverse events
nausea, vomiting, throat discomfort, bleeding, etc
Time frame: one week
number of malfunctions of portable endoscopy system
hydropower and gas failure, image display failure, disposable sheathed system damage, etc.
Time frame: 24 hours