Objectives: This study aims to determine if the device provides an advantage over traditional free hand puncture by comparing the number of attempts required to obtain access, time to achieve access, radiation exposure to patient and surgeon; and complications.
This is a single centre, participant randomized, open label study to evaluate the efficacy and safety of automated needle targeting (ANT-X) system compared to traditional free hand puncture for achieving renal access in Percutaneous lithotomy (PCNL) performed by urologists in training. Prior to performing kidney puncture on patients, all participating surgeons will undergo a simulation using Medical Simulation 12 balls phantom device with and without use of the ANT-X. Four urologic trainees (with similar little previous PCNL experience) will be recruited for the simulation. Two end points will be assessed: Number of attempts to achieve desired renal puncture and total fluoroscopy time. This simulation via 12 balls phantom device will involve fluoroscopic screening of calyx and successful puncture by urologic trainees under consultant supervision. Trainees will have 30 attempts of free hand puncture and 30 attempts using the ANT-X. The trainee will only move on to patient's renal puncture once they are able to achieve more than 80% of satisfactory renal puncture on simulator. For human renal puncture, we will aim for 60 cases of PCNL, 30 to be done using ANT-X puncture and 30 using traditional free hand puncture performed by 4 urologic trainees (15 cases per urologic trainee).
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
60
ANT-X device is a software-controlled automated electromechanical arm. It is developed to provide needle guidance for surgeon in MIS. This aims to reduce the surgery time taken and also minimize the radiation withstand by both surgeon and patient. ANT-X device is actually a Delta parallel robot implementing inverse kinematics to achieve effector's movement. Parallel structures excel in their great stiffness and high positioning accuracy. Most of the main components of ANT-X robot are made up of medical grade Polyether ether ketone (PEEK) material. PEEK is preferred to build the robot because it is radiolucent and lightweight while being able to provide relatively high material strength. The device is intended for use together with C-arm fluoroscope (fluoroscopy/CT technology). The fluoroscopy technology is used to provide image guidance for the needle
Number of attempts required to obtain access
An access attempt is defined as each separate passage of the needle into the kidney. Number of attempts by Investigator/surgeon will be recorded.
Time frame: Day 0 surgery
Time for the access
Time to successful access into the collecting system will be calculated in minutes starting from the placement of the needle on the skin surface to the identification of urine efflux from the end of the needle or reach of target. Start time of needle placement and time of urine efflux from the end of the needle or reach of target will be recorded in subject CRF in hh:mm:ss format.
Time frame: Day 0 surgery
Radiation exposure to patient and surgeon
Time frame: Day 0 surgery
Assessment of AEs
Advere events associated with the procedure
Time frame: Admission day till discharge within 7 days