The purpose of this study is to evaluate the safety and effectiveness of the T2Resistance Panel by validating clinical performance in three study arms: 1. Prospective arm: positive percent agreement (PPA) and negative percent agreement (NPA) against genetic determinants of resistance detected in whole blood clinical samples or isolates collected from positive blood cultures prospectively collected clinical samples of whole blood. 2. Contrived arm: positive percent agreement (PPA) and negative percent agreement (NPA) against samples with known status, via spiking healthy whole blood samples with bacterial strains harboring the resistance gene targets on the T2Resistance Panel. 3. Healthy donor arm: negative percent agreement (NPA) with presumed negativity of healthy donor whole blood samples. The data from all arms of the study will be used to support the Premarket Notification for the T2Resistance Panel to the U.S. Food and Drug Administration. Primary Endpoints The primary endpoints of this study with the T2Resistance Panel are estimated sensitivity, specificity, and safety.
The purpose of this study is to evaluate the safety and effectiveness of the T2Resistance Panel by validating clinical performance in three study arms: 1. Prospective arm: positive percent agreement (PPA) and negative percent agreement (NPA) against genetic determinants of resistance detected in whole blood clinical samples or isolates collected from positive blood cultures prospectively collected clinical samples of whole blood. 2. Contrived arm: positive percent agreement (PPA) and negative percent agreement (NPA) against samples with known status, via spiking healthy whole blood samples with bacterial strains harboring the resistance gene targets on the T2Resistance Panel. 3. Healthy donor arm: negative percent agreement (NPA) with presumed negativity of healthy donor whole blood samples. The data from all arms of the study will be used to support the Premarket Notification for the T2Resistance Panel to the U.S. Food and Drug Administration. Primary Endpoints The primary endpoints of this study with the T2Resistance Panel are estimated sensitivity, specificity, and safety. Sensitivity The estimated sensitivity of the T2Resistance Panel will be derived from two metrics: 1. Prospective arm: positive concordance between a positive result on the T2Resistance Panel and a positive sequencing result from a whole blood sample or isolate from positive blood culture 2. Contrived arm: positive concordance between a positive result on the T2Resistance Panel and a sample spiked with a bacterial isolates sequence confirmed to be harboring a known resistance gene on the T2Resistance Panel. The Contrived arm of the study will consist of ≥ 350 whole blood samples spiked with titer levels ranging from \< 1 CFU/mL to 100 CFU/mL. These samples will be provided to selected test sites for testing. Sensitivity values will be calculated separately for each channel and study arm, e.g., Prospective and Contrived study arms for blaKPC, blaCTX-M, blaNDM / blaVIM / blaIMP, blaOXA-48 Group, vanA / vanB, mecA / mecC and AmpC (blaCMY / blaDHA) channels. Specificity The estimated specificity of the T2Resistance Panel will be derived from three metrics: 1. Prospective arm: negative concordance between a negative ("Target not Detected") result on the T2Resistance Panel and a negative sequencing result from whole blood sample or isolate from positive blood culture 2. Healthy donor arm: negative concordance between a negative ("Target not Detected") result on the T2Resistance Panel and a presumed negativity of healthy donor whole blood sample 3. Contrived arm: negative concordance between a negative ("Target not Detected") result in a given channel of the T2Resistance Panel and the presence of a spiked bacteria known to not harbor the given resistance gene on the T2Resistance Panel Specificity values will be calculated separately for each channel and study arm, e.g., Prospective, Healthy donor, and Contrived study arms for blaKPC, blaCTX-M, blaNDM / blaVIM / blaIMP, blaOXA-48 Group, vanA / vanB, mecA / mecC and AmpC (blaCMY / blaDHA) channels. Safety There are no expected adverse events that are directly related to the T2Resistance Panel and the T2Dx Instrument as a result of participating in this study. Since the device does not come in contact with the patients and the test results are not used in clinical practice as part of standard of care, the only adverse events associated with study participation is the collection of blood samples for T2Resistance Panel testing.
Study Type
OBSERVATIONAL
Enrollment
1,566
The T2Resistance Panel is an in vitro diagnostic medical device that runs on the T2Dx instrument and detects bacterial markers that are commonly associated with antibiotic resistance. T2Resistance Panel detects thirteen (13) markers of resistance in seven detection channels: 1. blaKPC 2. blaCTX-M 3. blaNDM / blaVIM / blaIMP 4. blaOXA-48 Group 5. vanA / vanB 6. mecA / mecC 7. AmpC (blaCMY / blaDHA)
University of Alabama at Birmingham
Birmingham, Alabama, United States
Zuckerberg San Francisco General Hospital
San Francisco, California, United States
Tampa General Hospital
Tampa, Florida, United States
Ochsner Medical Center
New Orleans, Louisiana, United States
Johns Hopkins University
Baltimore, Maryland, United States
University Health, Truman Medical Center
Kansas City, Missouri, United States
New York University Langone Health
Brooklyn, New York, United States
University of Pittsburgh
Pittsburgh, Pennsylvania, United States
Rhode Island Hospital
Providence, Rhode Island, United States
University of Texas Health Science Center
Houston, Texas, United States
Sensitivity - Prospective
Positive concordance between positive T2Resistance panel results for positive samples.
Time frame: Six months from collection, maximum.
Specificity - Prospective
Negative concordance between a negative ("Target not Detected") result on the T2Resistance Panel and a negative sequencing result or a known negative sample.
Time frame: Six months from collection, maximum.
Safety / Adverse Events
Potential adverse events associated with the collection of blood samples for T2Resistance Panel Testing.
Time frame: During patient blood draw or immediately after blood draw, typically <1 hour.
Sensitivity - Contrived
Concordance between T2Resistance Panel results and positive blood culture, and contrived samples.
Time frame: Six months from collection, maximum.
Specificity - Contrived
Concordance between a negative ("Target not Detected") result on the T2Resistance Panel and sensitivity results associated with antibiotic susceptibility testing from positive blood culture samples or positive contrived samples.
Time frame: Six months from collection, maximum.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.