This is an in vivo 24-hour recovery study of leukoreduced red blood cells (RBCs) after automated separation of whole blood by the Reveos Automated Blood Processing System and storage for 42 days.
The Reveos® Automated Blood Processing System (Reveos system) is an automated whole blood processing system, not cleared for use in the United States. The Reveos System has been available outside the United States since 2012 and is seeking United States Food and Drug Administration (FDA) 510(k) clearance as an automated whole blood processor. The Reveos System is an integrated manufacturing system that processes whole blood (WB) units into blood components and is designed to minimize manual blood processing variables resulting from processing time, product and procedural variability. The Reveos System can simultaneously process up to 4 whole blood units into 4 RBC units, 4 leukoreduced plasma units, and 4 interim platelet units (IPU). The CPD/AS-5 RBC product is leukoreduced and may be stored up to 42 days. Terumo BCT is seeking 510(k) clearance from FDA/CBER for the Reveos® Automated Blood Processing System (Reveos System). The proposed study will provide validation on the in vivo performance of LR-RBC units derived from WB processed with the Reveos System. The results will provide data as to whether the Reveos System produced LR-RBC can maintain required levels of 24-hour in vivo recovery that meet FDA criteria for RBC.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
37
This is a non-interventional study, observing the in vivo performance of LR-RBC units derived from WB processed with the Reveos System.
Hoxworth Blood Center, University of Cincinnati
Cincinnati, Ohio, United States
Bloodworks Northwest Research Institute
Seattle, Washington, United States
Number of Participants With Success of RBC's in Vivo Recovery After 42-day Storage
The primary outcome of this study is to measure mean 24-hour, post-transfusion, in vivo RBC recovery percentage after LR-RBCs derived from WB and processed with the Reveos System have been stored for 42-days at 1-6°C. The 3 parts of FDA's RBC recovery criteria that must be met (below) are presented in Primary Outcome 1 and Primary Outcome 2. The derivation of the sample size was based on the level of precision attained with 24 subjects relative to the 1st part of the criteria. 1. The one-sided 95% lower confidence limit for the population proportion of success is greater than 70% where success of a unit is defined as the RBC in vivo 24-hour percentage recovery ≥ 75% \[Allows for low recoveries (\<75%) in 2/20 or 3/24 volunteers\] - Primary Outcome 1 2. The sample mean of percent recovery ≥ 75% - Primary Outcome 2 3. The sample standard deviation of in vivo 24-hour RBC recovery ≤ 9% - Primary Outcome 2
Time frame: 43 days
Mean 24-hour, Post-transfusion, in Vivo RBC Recovery Percentage After 42-day Storage
The primary outcome of this study is to measure mean 24-hour, post-transfusion, in vivo RBC recovery percentage after LR-RBCs derived from WB and processed with the Reveos System have been stored for 42-days at 1-6°C. The 3 parts of FDA's RBC recovery criteria that must be met (below) are presented in Primary Outcome 1 and Primary Outcome 2. The derivation of the sample size was based on the level of precision attained with 24 subjects relative to the 1st part of the criteria. 1. The one-sided 95% lower confidence limit for the population proportion of success is greater than 70% where success of a unit is defined as the RBC in vivo 24-hour percentage recovery ≥ 75% \[Allows for low recoveries (\<75%) in 2/20 or 3/24 volunteers\] - Primary Outcome 1 2. The sample mean of percent recovery ≥ 75% - Primary Outcome 2 3. The sample standard deviation of in vivo 24-hour RBC recovery ≤ 9% - Primary Outcome 2
Time frame: 43 days