Prospective cross-sectional diagnostic accuracy study with 1,000 patient participants and 30 lay provider/health care worker participants. The patient participant population will be recruited at clinics. At the point of care, study staff will collect capillary blood samples and conduct malaria microscopy (both the standard of care and study specific research-grade microscopy) and two index tests: the NxTek™ Malaria P.f plus Rapid Diagnostic Test (RDT) and the NxTek™ Malaria P.f/P.v RDT. Venous blood will be collected and transferred to the laboratory where the reference PCR assay and three comparator RDTs will be run. The index RDTs will also be repeated in the laboratory. All clinical management of study participants will follow the standard of care for malaria diagnosis in Brazil and will be based on the standard microscopy result. Confirmatory testing may also be conducted, such as typing and sequencing of Plasmodium genes and antigens of interest, including but not limited to HRP2, HRP3, and pLDH as well as the human inflammatory response marker CRP. The lay provider/health worker participants will include intended users of the index tests. They will be surveyed to assess index test usability through a questionnaire to assess label and packaging comprehension as well as results interpretation.
Primary Objectives 1.1 To assess the sensitivity, specificity and, when applicable, positive and negative predicting values (PPV and NPV) \[altogether referred to hereafter as "diagnostic accuracy"\] of NxTek™ Malaria Pf Plus in intended use settings for detecting P. falciparum infections in capillary whole blood samples collected prospectively from patients with symptoms suggestive of malaria. 1.2 To assess the diagnostic accuracy of NxTek™ Malaria Pf/Pv Plus in intended use settings for detecting P. falciparum and P. vivax infections in capillary whole blood samples collected prospectively from patients with symptoms suggestive of malaria. Secondary Objectives 2.1 To assess the diagnostic accuracy of the study comparator tests in intended use settings for detecting P. falciparum and P. vivax infections in venous whole blood samples collected prospectively from patients with symptoms suggestive of malaria. 1.3 To determine the frequency of P. falciparum infections containing HRP2 and/or HRP3 mutations and assess the impact of those on HRP2-based RDT diagnostic accuracy. 1.4 To assess the comprehension of the test packaging and labeling among intended users (trained lay providers and trained health care workers). 1.5 To assess the ability to read and interpret the test result outputs among intended users (trained lay providers and trained health care workers).
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
1,030
The NxTek (TM) Malaria Pf Plus, developed by Abbott, is a sensitive rapid chromatographic immunoassay for the qualitative detection of histidine-rich protein 2 (HRP2) and pLDH on a single test line of Pf malaria in human whole blood. This test is a lateral flow test for in vitro professional diagnostic use and intended as an aid to early diagnosis of Malaria infection in-patients with clinical symptoms. The NxTek (TM) Malaria Pf/Pv Plus, developed by Abbott, is a sensitive rapid chromatographic immunoassay for the qualitative detection of histidine-rich protein 2 (HRP2) and Plasmodium lactate dehydrogenase (pLDH) of Plasmodium falciparum (Pf) on one test line and of pLDH of Plasmodium vivax (Pv) malaria on a second test line in human whole blood. This test is a lateral flow test for in vitro professional diagnostic use and intended as an aid to early diagnosis of Malaria infection in-patients with clinical symptoms.
Centro de Pesquisa em Medicina Tropical de Rondônia (CEPEM)
Porto Velho, Rondônia, Brazil
RECRUITINGDiagnostic accuracy of NxTek™ Malaria Pf Plus for the detection of P. falciparum infections
1.1 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of NxTek™ Malaria Pf Plus for the detection of P. falciparum infections in patients with symptoms suggestive of malaria.
Time frame: 5 months
Diagnostic accuracy of of NxTek™ Malaria Pf/Pv Plus for the detection of P. falciparum and P.vivax infections
1.2 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of NxTek™ Malaria Pf/Pv Plus for the detection of P. falciparum and P.vivax infections in patients with symptoms suggestive of malaria.
Time frame: 5 months
Diagnostic accuracy of comparator tests for the detection of P. falciparum and, when relevant, P.vivax infections
2.1 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of comparator tests for the detection of P. falciparum and, when relevant, P.vivax infections in patients with symptoms suggestive of malaria.
Time frame: 5 months
Frequency of P. falciparum infections containing HRP2 and/or HRP3 mutations.
2.2 Frequency of P. falciparum infections containing HRP2 and/or HRP3 mutations.
Time frame: 5 months
Diagnostic accuracy of the index tests for the detection of P. falciparum infections with HRP2 and/or HRP3 deletions
2.3 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of the index tests for the detection of P. falciparum infections with HRP2 and/or HRP3 deletions in patients with symptoms suggestive of malaria.
Time frame: 5 months
Usability: label comprehension
2.4 Percent of end users who can accurately comprehend key messaging included in the investigational device packaging and labels.
Time frame: 1 week
Usability: results interpretation
2.5 Percent of end users who can accurately interpret the investigational device result output.
Time frame: 1 week