To generate preliminary safety and effectiveness data for brain-responsive neurostimulation of thalamocortical networks as an adjunctive therapy in reducing the frequency of generalized seizures in individuals 12 years of age or older with Lennox Gastaut Syndrome (LGS) who are refractory to antiseizure medications. The intent is to determine the feasibility and the optimal design of a subsequent pivotal study in order to expand the indication for use for the RNS System as a treatment for patients with medically intractable LGS.
This study is a prospective two-stage single-blind feasibility cross-over study designed to provide early safety and preliminary evidence of effectiveness for combined bilateral brain-responsive neurostimulation of thalamocortical networks for the treatment of generalized seizures in patients with LGS. Twenty participants will be treated with the RNS System across six Comprehensive Epilepsy Centers in the U.S. Enrollment will be staged in two cohorts of 10. Once all 10 participants of the first cohort complete Treatment Block 1 and the interim analysis criteria are met, the next cohort of 10 participants will be enrolled. The research study comprises six study periods: 1. Baseline Period 2. Implant (surgery) 3. Post-Op Period 4. Blinded Evaluation Period (which is made up of 3 treatment blocks, one of which is a sham stimulation) 5. Open Label Period 6. Long Term Follow-Up Period Two neurostimulators will be placed: one in the parieto-temporal skull on the left, and the second in the homologous region on the right. Depth leads will target the bilateral CM. Cortical strip or depth leads will target the prefrontal cortex. Each neurostimulator will be connected to two ipsilateral leads: one in the prefrontal cortex and one in the CM. A total of two neurostimulators and four leads will be implanted. During the Blinded Evaluation Period, the participant will move through 3 different treatment blocks (Treatment Block1, Treatment Block 2, and Treatment Block 3) in a random order. Two of the blocks will have active stimulation treatment and one of the blocks will have no (sham) stimulation treatment. The participant and caregiver will be blinded to the treatment condition. * Condition A - active treatment of high frequency short burst stimulation * Condition B - active treatment of low frequency long burst stimulation * Sham - no stimulation treatment After completing the Blinded Evaluation Period the participant will transition to the 1-year Open Label Period and have study appointments every 3 months. After completing the 1-year Open Label Period the participant will transition to the Long Term Follow-up Period. The Long-Term Follow-up Period may last up to 2 years with appointments every 3 months until the participant complete this research or this research study ends.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
DOUBLE
Enrollment
24
The RNS System provides closed loop responsive brain stimulation. The Neurostimulator monitors the electrical activity of the brain to detect abnormal activity that could lead to a seizure. If abnormal activity is detected, the neurostimulator delivers electrical stimulation to the brain through the leads to help prevent the seizure before it occurs.
University of Alabama at Birmingham
Birmingham, Alabama, United States
University of California, San Francisco
San Francisco, California, United States
Emory University
Atlanta, Georgia, United States
Massachusetts General Hospital
Boston, Massachusetts, United States
NYU Langone Medical Center
New York, New York, United States
Mount Sinai Hospital
New York, New York, United States
Safety: Annual device-related serious adverse event (SADE) rate
The lower limit of the 95% confidence interval of the annual SADE rate at months 12 and 24 post-implant is less than 33.6% and 22.0%, respectively (twice the SADE rate of 16.8% and 11.0%, respectively, at the same time points in the RNS System pivotal study)
Time frame: 12 months post-implant
Safety: Annual device-related serious adverse event (SADE) rate
The lower limit of the 95% confidence interval of the annual SADE rate at months 12 and 24 post-implant is less than 33.6% and 22.0%, respectively (twice the SADE rate of 16.8% and 11.0%, respectively, at the same time points in the RNS System pivotal study)
Time frame: 24 months post-implant
Effectiveness: Blinded evaluation period (BEP) responder rate
The responder rate during one or both of the stimulation conditions (A or B) is ≥ 30%. The responder rate is the proportion of participants that are responders. A responder in this study is defined as a participant who has a ≥ 35% reduction in the frequency of drop seizures compared to that participant's pre-implant baseline.
Time frame: 12 months post-implant
Safety: Post-op SAE rate
The lower limit of the 95% confidence interval of the SAE rate following the implant procedure (through 4 weeks post-op) is less than 24% (twice the SAE rate of 12% at the same time point in the RNS System pivotal study).
Time frame: 4 weeks post-implant