The objective of this clinical investigation is to evaluate the safety and technical success of the Basecamp Vascular controllable directional GECKO guidewire when used to facilitate endovascular access to the targeted vessel in order to treat the vascular lesion.
The Basecamp Vascular trial is a prospective monocenter, non-randomized, open clinical investigation of their controllable directional GECKO guidewire which has not yet received the CE mark. The GECKO active guidewire is a sterile, single-use device to be used by interventional neuroradiologists (INR) to position a distal access catheter (DAC). The device is a system composed of a deflector (the guide) and a handle. The handle contains the battery and integrated electronics. The device can be activated using buttons on the handle. When the user presses a button, a current is transmitted along the guide to its end: this allows the guide to bend (two possible curves). The end of the device can then take an "S" shape. In INR, the placement of a DAC (at the cervical level) is the first step of an endovascular intervention (Primary Access). This is followed by the treatment phase of the vascular pathology with other medical devices (example: embolization, angioplasty). The GECKO active guide is a system that facilitates endovascular navigation and allows the positioning of the DAC via vascular access (femoral or radial artery). The GECKO device is intended to introduce devices used for treatment, at a desired anatomical location (cervical carotid artery, for example) in the central circulatory system. The GECKO guidewire itself is not meant to perform a treatment but is used to set the primary access that will allow performance of the treatments.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
10
The GECKO device is intended to introduce devices used for treatment, at a desired anatomical location (cervical carotid artery, for example) in the central circulatory system. The GECKO guidewire itself is not meant to perform a treatment but is used to set the primary access that will allow performance of the treatments.
Fondation Ophtalmologique Adolphe de Rothschild
Paris, France
The cumulative incidence of all adverse events (serious and non-serious)
The cumulative incidence of adverse events (serious and non-serious) related to the non-permanent implanted guidewire device or the procedure occurring from Day 1 to 30 days post-procedure. All Serious Adverse Events will be reported (e.g., any thromboembolic event \[intracranial embolus in the targeted vessel territory or other territory, targeted vessel dissection or dissection of any catheterized vessel\], major or minor stroke, or death within 30 days of the intervention).
Time frame: 30 days
Number of successful GECKO guidewire placements at the target vessels
Technical success is defined as successful navigation, access and correct positioning of the guidewire and the distal access catheter at the target vessel. This will be measured by the number of successful GECKO guidewire placements at the target vessels.
Time frame: During procedure
Measurement of the "time"
Measurement of the "time" to navigate the distal access catheter for Primary Access into the target vessel (time from femoral or radial puncture until arrival at the target vessel).
Time frame: During procedure
GECKO Procedure Time
Defined as the time measured from the insertion of the GECKO guidewire until the removal of the GECKO guidewire.
Time frame: During procedure
Overall Procedure Time
Defined as the time measured from the insertion of the groin or radial access sheath/catheter until the removal of the access sheath
Time frame: During procedure
Number of participants with intracranial hemorrhage
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Number of participants with intracranial hemorrhage inclusive of subarachnoid, intraventricular or intraparenchymal hemorrhage (symptomatic).
Time frame: 30 Days
All-cause mortality.
All-cause mortality.
Time frame: 30 Days
Devices Used
All devices used will be recorded as successful or unsuccessful in conjunction with the GECKO guidewire.