This prospective imaging study is evaluating the feasibility of using the Halcyon 4.0 radiotherapy system for radiation therapy planning in patients with cancer. The Halcyon 4.0 system has been engineered to decrease the image acquisition time and the radiation exposure, but the system has not yet been clinically validated for use in radiation planning. This pilot study will evaluate images obtained on the Halcyon 4.0 system to assess if the quality is sufficient for radiation treatment plan construction.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DEVICE_FEASIBILITY
Masking
NONE
Enrollment
52
During each of the 1 or more imaging sessions, no more than 6 images total will be acquired. Only 5 images total across all of the imaging sessions will be used toward the study endpoint.
Washington University School of Medicine
St Louis, Missouri, United States
Percentage of images that are of sufficient quality for patient primary treatment plan construction
If the treatment target and surrounding organs at risk can be visualized and contoured for treatment planning purposes and a treatment plan created that meets the stated objectives ( or better than the stated objectives), then the kV CBCT images will be deemed feasible.
Time frame: At 9 weeks