The aim of the study is to compare functional results and complications of 3 methods of cataract phacoemulsification: bimanual 1.4 mm cataract surgery (B-MICS), coaxial 1.8 mm cataract surgery (C-MICS) and coaxial 2.4 mm small incision cataract surgery.
Reduction of the width of the corneal incision was one the main changes taking place in cataract surgery in recent years. The common use of foldable intraocular lenses (IOLs) and technological development of phaco machines allowed to reduce clear corneal incision below 3 mm. Term of Microincision Cataract Surgery (MICS) understood as cataract phacoemulsification performed with the incision width below 2 mm was defined by professor Alio in 2003. However, despite various modifications introduced in recent years, phacoemulsification still causes damage of the tissues that results in surgically induced astigmatism. Two MICS techniques have been developed: bimanual microincision cataract surgery (B-MICS) and coaxial microincision cataract surgery (C-MICS). In the bimanual technique cataract phacoemulsification can be performed through the main incision 1.4 mm wide due to the usage of sleeveless phaco tip (without irrigation) and irrigation chopper. The advantage of separation the irrigation from aspiration is improvement of liquid dynamics in the anterior chamber. Moreover, due to the usage of the irrigation chopper, in B-MICS it is possible to lower the mean ultrasound energy. In coaxial technique MICS phacoemulsification is performed through the incision 1.8 mm wide with usage of phaco tip with a silicon irrigation sleeve. The aim of the study is to compare functional results and complications of 3 methods of cataract phacoemulsification: bimanual 1.4 mm cataract surgery (B-MICS), coaxial 1.8 mm cataract surgery (C-MICS) and coaxial 2.4 mm small incision cataract surgery. Moreover, this study aimed to evaluate the impact of corneal width on best corrected visual acuity (uncorrected and corrected), surgically induced astigmatism, endothelial cell loss, intraocular pressure, anterior segment of the eye and central retinal thickness.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
90
microincision cataract surgery (phacoemulsification) with implantation of a foldable, acrylic, hydrophilic IOL Incise® MJ14, Bausch \& Lomb
microincision cataract surgery (phacoemulsification) with implantation of a foldable, acrylic, hydrophilic IOL Incise® MJ14, Bausch \& Lomb
small incision cataract surgery (phacoemulsification) with implantation of a foldable, acrylic, hydrophilic IOL Incise® MJ14, Bausch \& Lomb
the best corrected visual acuity
examined with digital Snellen chart, measured in decimal scale
Time frame: 36 months
the best uncorrected visual acuity
examined with digital Snellen chart, measured in decimal scale
Time frame: 36 months
autorefractometry
measured in diopters (D)
Time frame: 36 months
keratometry
measured in diopters (D)
Time frame: 36 months
intraocular pressure (IOP)
measured in millimeters of mercury (mmHg)
Time frame: 36 months
endothelial cell density (ECD)
measured in endothelial cells/mm2
Time frame: 36 months
central corneal thickness (CCT)
measured with Anterior Segment Optical Coherence Tomography (AS-OCT), unit of measure: µm (micrometers).
Time frame: 36 months
peripheral temporal corneal thickness
measured at the main incision site with Anterior Segment Optical Coherence Tomography (AS-OCT), unit of measure: µm (micrometers)
Time frame: 36 months
peripheral nasal corneal thickness
measured with Anterior Segment Optical Coherence Tomography (AS-OCT), unit of measure: µm (micrometers)
Time frame: 36 months
anterior chamber depth (ACD)
measured with Anterior Segment Optical Coherence Tomography (AS-OCT) in mm (milimeters).
Time frame: 36 months
the white-to-white (WTW) distance - the horizontal corneal diameter
measured with Anterior Segment Optical Coherence Tomography (AS-OCT, unit of measure: mm (milimeters).
Time frame: 36 months
length of the clear corneal incision
the chord length measured with Anterior Segment Optical Coherence Tomography (AS-OCT), unit of measure: µm (micrometers)
Time frame: 36 months
anterior chamber angle
measured with Anterior Segment Optical Coherence Tomography (AS-OCT) in horizontal scan, unit of measure: degrees
Time frame: 36 months
central foveal thickness (CFT)
measured with Optical Coherence Tomography (OCT), unit of measure: µm (micrometers)
Time frame: 36 months
parafoveal retinal thickness
measured with Optical Coherence Tomography (OCT), unit of measure: µm (micrometers)
Time frame: 36 months
surgically induced astigmatism SIA
measured in dioptres (D) - three different methods were used for calculation of surgically induced astigmatism (SIA): vector analysis method, vector decomposition method (C90), Naeser's polar values method (dKP-90)
Time frame: 36 months