A prospective study in paediatric, adolescent and young adult patients aged 7 to 18 years to evaluate the agreement between QTc measured using 12 lead electrocardiogram (ECG) and the wearable device ECG.
Rationale: Leveraging existing wearable technology has the ability to scale routine screening strategies across large populations and can facilitate ambulatory care. In recent years, wearable technology has rapidly diffused into consumer markets and provides unique opportunities to engage individuals on health issues of personal interest, and to collect personal health data. One example of such technology is the ability to perform electrocardiogram (ECG) type traces on wearable devices. However, utilization of this data source to improve patient outcomes has not been fully realized in the health care field. An app-based heart ECG algorithm that can be remotely analysed for QTc prolongation with potential to then notify the user of such potential irregularities could lead to widespread screening, detection and initiation of treatment solutions (i.e. correction of electrolytes, alteration of exacerbating medications, cardiac monitoring etc). Trial Design: This will be a prospective, single arm, experimental, non-significant risk study conducted with the assistance of eligible participants. Parents or mature minors will download the Beat to Beat app, once consented. Once consented the wearable device will be paired to the parent/guardians phone following the steps provided by the wearable device app. Once the device is paired and the study coordinator has assisted with the download of 'Beat to Beat' app, the patients will then be able to commence the ECG readings. A 12 lead ECG will be performed by the study team and immediately following this, an ECG will be recorded on the wearable device. The wearable device will be placed on the left wrist for a Lead I reading, on the left lower abdomen for a Lead II reading and in the fourth intercostal space left parasternal for a chest V2 Lead reading. Recording of the 12 lead ECG and wearable device ECG will be repeated again 4 days later. Once both timepoints (and two device recordings for each time period) have been completed the ECGs will be retrieved from the Beat to Beat app. Both the 12 lead and wearable device ECG will be de-identified and given a patient identifier before being uploaded into the REDCap database.
Study Type
INTERVENTIONAL
Allocation
NA
The app together with the wearable device will record a V1 (Left wrist) or V2 (left ankle) ECG reading. These measurements will be conducted on Day 1 and Day 4.
The 12 lead ECG will measure corrected QT interval by recording RR and QT intervals on Day 1 and Day 4. This will be performed within the same episode of care as the wearable device ECG recording. There is no need to perform both at exactly the same time point, but within a 15 minute window.
The Royal Children's Hospital
Parkville, Victoria, Australia
Recording of a 12 lead ECG and wearable device ECG
12 lead ECG and wearable device ECG recordings will be taken at the time point of care. The wearable device ECG will be placed on the left wrist to record a V1 (lead I) ECG and on the left ankle to record a V2 (lead II) ECG reading from which the QT interval can be calculated.
Time frame: Day 1 of inpatient stay
Recording of a 12 lead ECG and wearable device ECG
12 lead ECG and wearable device ECG recordings will be taken at the time point of care. The wearable device ECG will be placed on the left wrist to record a V1 (lead I) ECG and on the left ankle to record a V2 (lead II) ECG reading from which the QT interval can be calculated.
Time frame: Day 4 of inpatient stay
Calculation of QT interval by two blinded health professionals
The wearable device ECG and 12 lead ECG will be de-identified and given appropriate study numbers. This data will be listed separately for each patient. The QT interval for each recording will be calculated. This is the time corresponding to beginning of depolarization to repolarization of the ventricles. It is calculated using a standardised approach (Fridericia's formula:QTcF = QT divided by cube root of RR).
Time frame: Day 1 of inpatient stay
Calculation of QT interval by two blinded health professionals
The wearable device ECG and 12 lead ECG will be de-identified and given appropriate study numbers. This data will be listed separately for each patient. The QT interval for each recording will be calculated. This is the time corresponding to beginning of depolarization to repolarization of the ventricles. It is calculated using a standardised approach (Fridericia's formula:QTcF = QT divided by cube root of RR).
Time frame: Day 4 of inpatient stay
Number of Participants with Abnormal QTc that is greater than 0.48mm on 12 lead ECG and wearable device ECG
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
45
From the analysis of QT measurements from Outcome 2, prolonged or abnormal QTc measurement will be noted. An abnormal QTc is greater than 0.48mm. This data will be collected separately for each patient.
Time frame: Day 1 inpatient stay
Number of Participants with Abnormal QTc that is greater than 0.48mm on 12 lead ECG and wearable device ECG
From the analysis of QT measurements from Outcome 2, prolonged or abnormal QTc measurement will be noted. An abnormal QTc is greater than 0.48mm. This data will be collected separately for each patient.
Time frame: Day 4 inpatient stay
Sensitivity calculations of wearable device vs 12 Lead ECG
Sensitivity is calculated only in the participants who are declared to have QTc prolongation (abnormal QTc) and is calculated as the proportion patients with QTc prolongation on the wearable device among those with QTc prolongation on the 12-lead ECG.
Time frame: Day 1 inpatient stay
Sensitivity calculations of wearable device vs 12 Lead ECG
Sensitivity is calculated only in the participants who are declared to have QTc prolongation (abnormal QTc) and is calculated as the proportion patients with QTc prolongation on the wearable device among those with QTc prolongation on the 12-lead ECG.
Time frame: Day 4 inpatient stay
Specificity calculations of wearable device vs 12 Lead ECG
Specificity is calculated only in participants who do not have QTc prolongation and is calculated as the proportion patients who do not have QTc prolongation on the wearable device among those who do not have QTc prolongation on the 12-lead ECG.
Time frame: Day 1 inpatient stay
Specificity calculations of wearable device vs 12 Lead ECG
Specificity is calculated only in participants who do not have QTc prolongation and is calculated as the proportion patients who do not have QTc prolongation on the wearable device among those who do not have QTc prolongation on the 12-lead ECG.
Time frame: Day 4 inpatient stay
To calculate the interobserver variability between the two health care professional readings of QTc.
The corrected QT interval will be calculated on both the wearable device ECG and the 12 lead ECG for each patient and measured by a health professional. QT interval will be calculated by the health professional by using the QT and RR interval from the ECGs using Frederichia's interval (QTcF = QT divided by cube root of RR). This data will be collected separately for each patient. The standard deviation of the difference in QTc interval between the two observers will be reported.
Time frame: Day 1 inpatient stay
To calculate the interobserver variability between the two health care professional readings of QTc.
The corrected QT interval will be calculated on both the wearable device ECG and the 12 lead ECG for each patient and measured by a health professional. QT interval will be calculated by the health professional by using the QT and RR interval from the ECGs using Frederichia's interval (QTcF = QT divided by cube root of RR). This data will be collected separately for each patient for each patient The standard deviation of the difference in QTc interval between the two observers will be reported.
Time frame: Day 4 inpatient stay