The study will evaluate the mean average error (MAE) of measuring pulse rate using the Norbert Device (ND) as compared to a reference heart rate measured using an electrocardiogram (ECG).
The study design is to validate the pulse and heart rate accuracy as required per ISO 80601-2-61, Clause 201.12.1.104. The study is a non-randomized, non-blinded single arm design to evaluate the clinical accuracy of the ND's pulse rate measurement compared to a reference heart rate manually measured by observing R-peaks in a standard concurrently recorded ECG from an FDA 510(k)-cleared reference clinical ECG (RCECG). The sponsor will ensure that the study staff enrolls a range of types of arrhythmias including tachycardia, bradycardia, atrial fibrillation, bigeminy, trigeminy, and others.
Study Type
OBSERVATIONAL
Enrollment
37
The Norbert Device (ND) is a contactless, noninvasive device that provides measurements of body temperature, oxygen saturation (SpO2), and pulse rate.
Newlab
Brooklyn, New York, United States
Staten Island University Hospital
Staten Island, New York, United States
Mean Average Error (MAE)
MAE will be evaluated by Mean absolute error (MAE) of the pulse rate of the device compared to the heart rate measured from the RCECG.
Time frame: Up to 1 hour