Accurate, precise, and comparable hemoglobin measurements is of great importance, both for clinical value in diagnosing anemia and ensuring pregnant women receive appropriate treatment. The Masimo Total Hemoglobin SpHb® is a continuous and non-invasive handheld device with an optical sensor placed on the finger that measures hemoglobin levels using pulse oximetry. The objective of this study is evaluate the compatibility of hemoglobin measurements between SpHb and the gold standard laboratory-based assessment (complete blood count assessed via five-part autoanalyzer) throughout the course of pregnancy and at six weeks postpartum.
Anemia, a condition that affects an estimated 613 million (33%) women of reproductive age worldwide and is classified as a moderate to severe public health problem in many countries, is a secondary outcome in the PRiSMA MNH study. Three current PRiSMA MNH study sites conduct non-invasive and continuous hemoglobin monitoring with a Masimo device (Kenya, Pakistan, Zambia). Accurate, precise, and comparable hemoglobin measurements is of great importance, both for clinical value in diagnosing anemia and ensuring pregnant women receive appropriate treatment. The Masimo Total Hemoglobin SpHb® is a continuous and non-invasive handheld device with an optical sensor placed on the finger that measures hemoglobin levels using pulse oximetry. The measurement takes under one minute and does not require blood samples or laboratory testing. These characteristics make it a particularly promising medical technology for low-resource areas. Previous research studies have found that SpHb is able to accurately detect hemoglobin levels in adult patients with a similar degree of bias and standard deviation to point-of-care (PoC) invasive method (e.g., HemoCue device) measurements. However, high variability in bias and in limits of agreements for the Masimo device was found in a study involving pregnant patients. The objective of this study is evaluate the compatibility of hemoglobin measurements between SpHb and CBC assessed via 5-part autoanalyzer throughout pregnancy and at six weeks postpartum. This study is nested in the Pregnancy Risk Stratification Innovation and Measurement Alliance (PRiSMA) Maternal and Newborn Health (MNH) study. PRISMA MNH is a population-based, open-cohort study that seeks to evaluate pregnancy risk factors and their associations with adverse pregnancy outcomes in five countries in sub-Saharan Africa and Southeast Asia. Three study sites (located in Zambia, Kenya, and Pakistan) will participate and collect hemoglobin data at five time points (\<20 weeks, 20 weeks, 28 weeks, 36 weeks gestation, and six weeks postpartum). The investigators will measure hemoglobin using a venous blood sample via auto-analyzer (gold standard) and the non-invasive device. The investigators will assess agreement between Masimo total hemoglobin and complete blood count and on a continuous scale using Intraclass Correlation Coefficient and Bland-Altman Analysis. The second objective is to assess agreement between the two measures on a binary scale using Positive Percentage Agreement and Negative Percentage Agreement, Cohen's Kappa, and McNemar Test. On an ordinal scale, agreement will be measured using Weighted Cohen's Kappa and Harrel's Concordance Index. Lastly, the investigators will assess factors that might affect the accuracy of Masimo total hemoglobin using linear mixed models.
Study Type
OBSERVATIONAL
Enrollment
900
Masimo Radical-67® Pulse CO-Oximeter and rainbow® sensors.
The test will be done using venous blood collected in EDTA tubes. Samples will be analyzed within 6 hours of collection or stored at 2 to 8C and analyzed within 24 hours. Analysis will be done using a five-part differential hematology analyzer.
Kenya Medical Research Institute-Center for Global Health Research
Kisumu, Kenya
Aga Khan University
Karachi, Pakistan
Kamwala District Health Centre
Lusaka, Zambia
Women and Newborn Hospital of the University Teaching Hospitals
Lusaka, Zambia
Hemoglobin (<20 weeks gestation)
Hemoglobin levels will be assessed using the Masimo device and via complete blood count (5 part autoanalyzer) during pregnancy: \<20 weeks gestation
Time frame: <20 weeks gestational age
Hemoglobin (20 weeks gestation)
Hemoglobin levels will be assessed using the Masimo device and via complete blood count (5 part autoanalyzer) during pregnancy: 20 weeks
Time frame: 18-25 weeks gestational age
Hemoglobin (28 weeks gestation)
Hemoglobin levels will be assessed using the Masimo device and via complete blood count (5 part autoanalyzer) during pregnancy: 28 weeks
Time frame: 26-30 weeks gestational age
Hemoglobin (36 weeks gestation)
Hemoglobin levels will be assessed using the Masimo device and via complete blood count (5 part autoanalyzer) during pregnancy: 36 weeks.
Time frame: 34 weeks gestational age until delivery
Hemoglobin (6 weeks postpartum)
Hemoglobin levels will be assessed using the Masimo device and via complete blood count (5 part autoanalyzer) at 6 weeks postpartum
Time frame: 6-12 weeks postpartum