A pulse oximeter is used for continuous monitoring of peripheral oxygen saturation (SpO2) and plays an important role in the early detection of hypoxia, guiding the titration of supplemental oxygen, and reducing the need for blood gas analysis. Therefore, the SpO2 accuracy of a pulse oximeter should be validated for patient safety. The objective of this study is to evaluate whether the CART-I pulse oximeter with a ring-type wearable PPG sensor (Sky Labs Inc., Seoul, Korea) provides clinically reliable SpO2 readings over the range of 70-100% SaO2, during steady-state, non-motion conditions.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
13
CART-I noninvasively measures the oxygen saturation of blood in a local area by screening the bloodstream through the user's finger using PPG signals
Sky Labs
Gyeonggi-do, Pangyo, South Korea
Accuracy of SpO2
Mean bias, standard deviation, standard error, maximum/minimum bias, precision are calcuated
Time frame: 1 day