The purpose of this study is to evaluate the accuracy and consistency of the X-Trodes acquisition system compared to a Food and Drug Administration (FDA)-cleared clinical electrophysiology device.
The study aims to demonstrate that the performance of the study X-Trodes acquisition system is equivalent to that of an FDA-cleared clinical electrophysiology devices (in measuring electroencephalography , electrooculography, electromyography, and electrocardiogram signals). Subjects will wear both devices simultaneously while signals are captured in resting state. Following, an evaluation of the similarity of signals between the two devices will be conducted. The evaluation will determine if the quality of the X-Trodes system matches that of the FDA-approved device.
Study Type
OBSERVATIONAL
Enrollment
57
No intervention. The participants will sit passively while electrophysiology signals are captured from two devices in order to evaluate the signal quality.
B-Cube
Chicago, Illinois, United States
Signals alignment between two devices
Electrophysiological time-series (EEG, EOG, ECG, EMG) in 2 recorded devices will be compared. For each subject, the time-series from 2 devices will be assessed by three independent raters (US board-certified neurologist/cardiologist). The raters will rank the quality of signals (1-signal is corrupt or unusable; 5-signal is clean and useable). Raters will be blind to the signal device. Time-series will be deemed interpretable if at least 2 of the raters scored it 4-5. Following the rater evaluation, the outcome will also be reported as a binary output where scores of 4-5 are considered interpretable signals. The trial will be considered successful if the number of high-quality (mean score 4-5) X-Trode series will be within one-standard deviation from the FDA-approved device's series while both systems yield over 80% high-quality signals.
Time frame: Within 3 months from the completion of data collection.
Assessment of agreement among raters
Agreement among raters will be evaluated by calculating the Intraclass correlation coefficient (ICC). ICC will be calculated for each of the 4 electrophysiological modalities. The acceptance criterion for ICC is 0.60 and higher. If no agreement among rates is observed (ICC \< 0.6) then Spearman correlation coefficient between each pair of raters for the specific signal (per subect, per time-series) will be calculated and the rankings of the two raters with the highest correlation coefficient will be selected for the analysis.
Time frame: Within 3 months from the completion of data collection.