This is a prospective and multicenter clinical investigation aiming to evaluate the safety and effectiveness of coronary crossing system for the treatment of patients with coronary chronic total occlusions.
This study is divided into two phases. The first phase is the pilot study stage, which is a prospective single-group observational test (FIM phase). The second phase is the pivotal study phase, which is a prospective, multi-center, single-group clinical trial (OPC phase). It is carried out in about 20 hospitals in China. 1. Pilot study phase (FIM phase): It is a prospective single-group observational test. It is expected to recruit 10 subjects in a research center in China. 2. Pivotal study phase (OPC phase): It is a prospective, multicenter, single-group clinical trial. 146 subjects are expected to be recruited in 20 research centers in China. 3. All subjects with coronary chronic total occlusions participating in this clinical study must have only a CTO lesion which occluded in length of more than 5 mm and located in a native coronary artery with a diameter of ≥ 2.50 mm (visual inspection). 4. All subjects receive clinical follow-up during hospitalization, 30 days after surgery.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
156
The mechanical vibration generated by the Coronary Crossing System through the piezoelectric effect is amplified and transmitted to the distal end of the guidewire to achieve high-frequency vibration at the tip of the guidewire, and then achieve efficient and safe penetration of the fibrous cap of the CTO.
Zhongshan Hospital of Fudan University
Shanghai, China
RECRUITINGDevice success
Device success is defined as successful antegrade crossing of the CTO into the true arterial lumen distal to the occlusion following advancement of at least 3 mm of the Coronary Crossing System into any segment as assessed by the core angiographic laboratory. (An antegrade only approach may involve Antegrade Dissection Re-Entry (ADR) with or without the use of specialized devices (e.g. Stingray); absence of successful retrograde crossing of a collateral will still be considered an antegrade only approach.)
Time frame: Baseline procedure
Overall device success
Overall device success is defined as successful crossing of the CTO, following use of the coronary crossing system during the procedure including antegrade or retrograde approach.
Time frame: Baseline procedure
Technical Success
Technical success is defined as successful antegrade or retrograde crossing of the CTO into the true arterial lumen following advancement of at least 3 mm of the Coronary Crossing System into the proximal segment, with \<30% residual stenosis of the CTO after successful treatment with drug-eluting stents or drug-eluting balloons, with antegrade flow≥TIMI grade 2 in all branch vessels ≥ 2.5 mm in diameter distal to the CTO, as assessed by the core angiographic laboratory.
Time frame: Baseline procedure
Procedural Success
Procedural success is defined as the achievement of Technical Success with freedom from device related MAEs as assessed by the Clinical Events Committee (CEC).
Time frame: Baseline procedure
Clinical Success
Clinical success is defined as the achievement of Procedural Success with freedom from device related MAEs through 48 hours or hospital discharge and 30 days post index procedure as assessed by the Clinical Events Committee (CEC).
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: 30 days post-procedure
Freedom from device related major adverse event(MAE)
Freedom from device related major adverse events (MAEs) at 48 hours or hospital discharge (whichever occurs first) post procedure, as assessed by Clinical Events Committee (CEC), defined as: * Cardiovascular related deaths * Myocardial Infarction. * Coronary Artery Perforation requiring treatment * Unscheduled cardiac surgery anytime between enrollment to completion of the study. * Intra-procedure stroke * Radiation exposure ≥ 8 Gy
Time frame: 48 hours or until discharge
Freedom from major adverse event(MAE)
Freedom from major adverse event(MAE) is defined as the freedom from device related MAE assessed through 30 days post-index procedure, as assessed by the Clinical Events Committee(CEC).
Time frame: 30 days post-procedure