The goal of this diagnostic study is to validate estimation of inspiratory muscle pressure by an artificial intelligence algorithm compared to the gold standard, the measure from an esophageal catheter balloon, in patients under assisted mechanical ventilation. The main questions it aims to answer are: • Are inspiratory muscle pressure estimates from an artificial intelligence algorithm accurate when compared to the direct measure from an esophageal balloon? Participants will be monitored with an esophageal balloon and with an artificial intelligence algorithm simultaneously, with inspiratory muscle pressure estimation during assisted mechanical ventilation with decremental levels of pressure support.
This is a diagnostic study to validate estimation of inspiratory muscle pressure during assisted ventilation from an artificial intelligence algorithm integrated in a mechanical ventilator (FlexiMag, Magnamed, Brazil) compared to direct measure of muscle pressure from esophageal catheter balloon (gold standard). This is a novel non-invasive method to estimate inspiratory muscle pressure. After obtaining informed consent, participants will be monitored simultaneously with the esophageal balloon and the artificial intelligence algorithm, with decremental levels of pressure support (20 to 2 cmH2O, in steps of 20 minutes). Esophageal balloon will be removed after completing the last pressure support step. The investigators estimated a sample of 50 participants, considering 3 cmH2O as a clinically relevant discordance between methods and 10% of missing data. Concordance analysis and correlation analysis will be performed. Procedures will follow a specific Standard Operating Procedures and participants inclusion data will be inserted in REDCap.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
50
Estimation of inspiratory muscle pressure by an artificial intelligence algorithm integrated in the mechanical ventilator (FlexiMag, Magnamed, Brazil).
Heart Institute, University of São Paulo
São Paulo, São Paulo, Brazil
Concordance between muscle pressure amplitude (in cmH2O) estimation by artificial intelligence and esophageal balloon
Analysis of the bias and limits of agreement (Bland-Altman plot) between muscle pressure estimated amplitude in cmH2O from artificial intelligence and measured by esophageal balloon.
Time frame: 4 hours
Correlation between muscle pressure amplitude estimation (in cmH2O) by artificial intelligence and esophageal balloon
Correlation, reported as R-squared and a correlation plot, between amplitude in cmH2O of muscle pressure estimation by artificial intelligence and esophageal balloon.
Time frame: 4 hours
Detection of initiation time and ending time of a spontaneous breathing cycle by artificial intelligence compared with esophageal balloon
Time difference (in ms) between initiation of a spontaneous breathing cycle and ending of a spontaneous breathing cycle between artificial intelligence and esophageal balloon.
Time frame: 4 hours
Sensitivity and specificity of patient-ventilator asynchrony automated detection using the Artificial Intelligence Muscle Pressure estimator
Number of patient-ventilator asynchronies detected using artificial intelligence compared with number of asynchronies detected by experts assessing airway pressure, flow and esophageal balloon waveforms.
Time frame: 4 hours