Compare the data obtained from the Raman analyzer and paraffin pathology examination on the same external brain tissue sample. Evaluate the effectiveness and safety of the Raman analyzer for intraoperative diagnosis gliomas of brain resection tissue samples, using paraffin pathological examination results as clinical reference standards.
Based on statistical calculations,108 positive samples and 148 negative samples will be included in the trial in all trial centers. Compare the results between the Raman analyzer and the paraffin pathological. And calculate the sensitivity, the specificity, and other indicators of Raman analyzer. During surgery, core or marginal tissue samples were taken from subjects. The test samples size:0.2cm\<length diameter ≤ 2cm. The sample testing result is based on the Raman test points. Then take the same tissue sample for paraffin pathological diagnosis. Statistical description of all data, including baseline data, all efficacy indicators, and all safety data. The measurement data give the mean, standard deviation, minimum, maximum, median,25 quantile and 75 quantile; Provide frequency and composition ratio for counting data. The baseline data was analyzed using the Full Analysis Set (FAS); The effectiveness analysis adopts FAS and PPS; The security analysis uses the Security Dataset (SS).
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
256
Perform two diagnostic methods on the same sample
Sensitivity of the Raman analyzer in detecting gliomar
Among the samples determined by paraffin pathology as gliomas, the percentage of samples detected by Raman analyzer as gliomas.
Time frame: Through study completion, an average of 1 year
Specificity of the Raman analyzer in detecting gliomar
Among the samples determined by paraffin pathology as non gliomas, the percentage of samples detected by the Raman analyzer as non gliomas.
Time frame: Through study completion, an average of 1 year
Using the paraffin test results as a reference, calculate the accuracy of the Raman analyzer in detecting gliomar
The proportion of tissue samples with consistent results between Raman analyzer detection and paraffin pathological diagnosis.
Time frame: Through study completion, an average of 1 year
Positive predictive value of the Raman analyzer in detecting gliomar
The percentage of samples diagnosed with glioma by paraffin pathology in the samples detected as glioma by Raman analyzer
Time frame: Through study completion, an average of 1 year
Negative predictive value of Raman analyzer in detecting gliomar
The percentage of samples diagnosed with non glioma by paraffin pathology in the samples detected as non glioma by Raman analyzer
Time frame: Through study completion, an average of 1 year
Kappa coefficient
Kappa coefficient≥0.75 indicates high consistency; 0.75\>Kappa coefficient≥0.4, considered consistent; If the Kappa coefficient is less than 0.4, it is considered inconsistent.
Time frame: Through study completion, an average of 1 year
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
The Raman analyzer usability evaluation
Use Evaluation Form
Time frame: Through study completion, an average of 1 year
Time consumption for the Raman analyzer in detecting gliomar
Time required from emitting laser to completing single point detection
Time frame: Through study completion, an average of 1 year