The purpose of this study is to validate cardiac output and stroke volume derived from Vitalstream against Gold Standard measurements obtained using thermodilution. The Vitalstream device is a continuous noninvasive physiological monitor (Caretaker Medical LLC, Charlottesville, Virginia, further referred to as "CTM") provides heart rate, continuous noninvasive blood pressure (BP), respiratory rate, stroke volume and cardiac output.
The device is currently undergoing IEC 60601 Medical Safety Standards testing and has already passed all FDA-required emission and discharge testing. The Vitalstream tracks central aortic BP via pulse analysis, specifically Pulse Decomposition Analysis ("PDA"), of the peripheral pulse at a distal site, typically finger. The device uses a low pressure \[30-40 mmHg\], pump-inflated, finger cuff that pneumatically couples arterial pulsations via a pressure line to a custom-designed piezo-electric pressure sensor for detection and analysis. Physiological data are communicated wirelessly to a tablet-based user interface via Bluetooth. The Vitalstream monitor is most beneficial as it is non-invasive, is a minimal risk device, and for this protocol will not be utilized to make treatment decisions for the study subject.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
58
The device will be placed on the subject and monitoring started in the holding room on the morning of surgery. Monitoring will continue postoperatively until the pulmonary artery catheter (PAC) is removed (Standard of Care) or discharged, or at study team discretion. At Intensive Care Unit (ICU) admittance and again 3 hours later a straight leg raise will be done to monitor for changes in hemodynamics.
Atrium Health Wake Forest Baptist
Winston-Salem, North Carolina, United States
Cardiac output values
Cardiac output values correlation of cardiac output between the PA catheter as assessed by the CCO and the VitalStream - the amount of blood pumped by the heart minute - Cardiac output is calculated by multiplying stroke volume with heart rate - Cardiac output in humans is generally 5-6 L/min in an at-rest to more than 35 L/min in elite athletes during exercise. HR is determined by signals from the sinoatrial node, which automatically depolarizes at an intrinsic rate of 60 to 100 times each minute - Low-output symptoms are caused by the inability of the heart to generate enough cardiac output. This leads to reduced blood flow to the brain and other vital organs
Time frame: hour 24
Blood pressure values
correlation of blood pressure continuously being monitored by the Vitalstream and the subjects indwelling arterial line
Time frame: hour 24