This study adapted a prospective, multicenter single-group objective performance criteria design. A total of 163 patients with non-valvular atrial fibrillation were enrolled in the clinical trial,follow up were scheduled in 7 days, 45 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5years after procedure.
This study adapted a prospective, multicenter single-group objective performance criteria design。A total of 163 patients with non-valvular atrial fibrillation were enrolled in the clinical trial, the patients signed the informed consent form and met the inclusion and exclusion criteria, the subjects were implanted with the SeaLA™ LAA (left atrial appendage) occlusion, and TTE(transthoracic echocardiography) examination was performed 7 days after procedure or before discharge (whichever comes first), TEE(transesophageal echocardiography) examination were performed 45 days and 6 months after procedure, and TTE examination at 12 months. In the 2nd, 3rd, 4th, and 5th years of telephone follow-up subjects, the main adverse event data were collected and analyzed, and the long-term safety and efficacy of the left atrial appendage closure device product after implantation were evaluated.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
163
1. Guided by esophageal ultrasound (TEE) and angiography, the diameter of the left atrial appendage anchoring zone and the longest depth of the left atrial appendage were measured to determine the appropriate size of the left atrial appendage occluder. 2. Maintain Activated Clotting Time (ACT) greater than 250 seconds after atrial septal puncture to the completion of the entire procedure. 3. Before the plugging device is released, confirm whether there is a residual peri-device leak, and if there is a residual peri-device leak, consider retracting the occluder to reposition the release or retract the occluder to choose a larger size occluder to re-occlusion. If ultrasound confirms residual peri-device leak greater than 5 mm, do not release and implant the left atrial appendage occluder. 4. Before and after release, confirm the position and stability of the occluder under ultrasound (TEE), and confirm that the occluder completely occludes the left atrial appendage.
General Hospital of the Northern Theater of the Chinese People's Liberation Army
Shengyang, China
Left atrial appendage closure rate
After the closure device is implanted, it is confirmed by ultrasound that no forward or reverse blood flow passes through the closure device, and the residual forward or reverse blood flow at the edge of the closure device is ≤ 3mm
Time frame: 6 months after procedure
Device success rate
smooth delivery of left atrial appendage occluder to left atrial appendage, successful release of left atrial appendage occlusion, smooth withdrawal of delivery system
Time frame: immediately after the procedure
Ischemic stroke rate
Ischemic stroke refers to the loss of nerve function caused by transient ischemic attack, cerebral thrombosis or cerebral embolism caused by the brain itself and/or systemic blood circulation disorders leading to cerebral blood supply disorders. Ischemic stroke is defined as a modified Rankin score (mRS) ≥2 90 days after stroke onset. The scale runs from 0 to 5, running from perfect health without symptoms to death, higher scores mean a worse outcome, 0 means no symptoms, 5 means dead.
Time frame: 12 months after the procedure
Major adverse events rate
Unexplained death or any device- or procedure-related complications, including but not limited to: death, severe cardiac tamponade requiring pericardiocentesis or surgery, instrument embolism, surgery-related stroke.
Time frame: 7 days after procedure