The goal of this clinical trial is to evaluate the safety, effectiveness, and performance of Venus-Neo Surgical Aortic Valve in subjects who are clinically indicated for aortic valve replacement.
This is a prospective, multicenter, non-randomized interventional study to evaluate the safety, effectiveness, and performance of Venus-Neo Surgical Aortic Valve. 155 subjects are estimated to be enrolled in this study. Total enrollment period for this trial is estimated to be 1 year. Follow-up duration is estimated to be 5 years. Overall duration of the trial is estimated to be 6 years. The trial begins with the enrollment of the first subject and ends after the last subject is exited from the trial after completing the last follow-up visit at approximately 5 years, all subjects are fully monitored, all outstanding data queries are resolved.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
155
Implant of a Venus-Neo Surgical Aortic Valve
Deutsches Herzzentrum Berlin-Charité
Berlin, Germany
primary composite safety endpoint
composite safety endpoint at 1 year, including: valve-related mortality, thromboembolism, aortic valve reintervention, structural valve deterioration, major paravalvular leak (defined as moderate or greater paravalvular leak or any paravalvular leak requiring intervention), valve thrombosis, endocarditis
Time frame: 1 year
secondary composite safety endpoint
safety is assessed over 1-year timeframe by comparing the occurrence of specific safety endpoints to the objective performance criteria (OPC) reported in Table I.1 in "ISO:5840-2:2021(E), Annex I, Methods of evaluating clinical data against objective performance criteria"
Time frame: 2 through 5 Years
Occurrence of each of the following adverse events
Occurrence of each of the following adverse events until 5 years, including: mortality, valve-related mortality, thromboembolism, aortic valve reintervention, structural valve deterioration, non-structural valve deterioration, all paravalvular leak, major paravalvular leak, valve thrombosis, endocarditis, all hemorrhage, major hemorrhage, bioprosthetic valve explant, hemolysis
Time frame: baseline, pre-discharge, 30 days, 6 months, 12 months and annually until 5 years thereafter