Prospective, randomized controlled, single-blind, multicenter, clinical trial to demonstrate the safety and efficacy of the SELUTION SLR 014 PTCA DEB for treatment of de novo lesions in small coronary vessels, defined as reference vessel diameter (RVD) of 2.00 mm to 2.75 mm, in support of a pre-market approval (PMA) application to the United States (US) FDA. The Study will enroll up to 910 randomized subjects, up to 30 subjects in a parallel angiographic substudy, and up to 20 subjects in a parallel pharmacokinetic (pK) substudy, at up to 80 sites in the US, Canada, Brazil, Japan and Europe. A minimum of 50% of the subjects will be enrolled in the US.
Prospective, randomized controlled, single-blind, multicenter, clinical trial The study will enroll up to 910 randomized subjects, up to 30 subjects in a parallel angiographic substudy, and up to 20 subjects in a parallel pharmacokinetic (pK) substudy, at up to 80 sites in the US, Canada, Brazil, Japan and Europe. A minimum of 50% of the subjects will be enrolled in the US. Subjects who present with chronic coronary syndrome (CCS), unstable angina or stabilized non-ST elevation myocardial infarction (NSTEMI) with an indication for percutaneous coronary intervention (PCI) with planned intervention for de novo lesions in small coronary vessels (RVD 2.00 mm to 2.75 mm) and meeting all eligibility criteria will be randomized 1:1 to treatment of the identified target lesion with either the SELUTION SLR 014 PTCA DEB or contemporary DES. Randomized Cohort: * Intervention (DEB Strategy): Subjects randomized to the SELUTION SLR 014 PTCA DEB arm will receive lesion preparation according to the 3rd drug coated balloon (DCB) consensus (optimal balloon angioplasty with adjunct treatment using high-pressure balloon, intravascular lithotripsy, laser, rotational or orbital atherectomy, cutting or scoring balloon at the discretion of the operator as needed to reduce diameter stenosis to ≤ 30%). Subjects with lesions that are then best treated by provisional stenting (flow-limiting dissection, residual stenosis \> 30%) will receive a DES instead of the SELUTION DEB but remain in the SELUTION DEB group (intention to treat analysis). The DEB should not be used after DES implant. * Control (DES): Subjects randomized to the DES arm will receive treatment using any FDA approved "limus-based" DES, as per standard institutional practice. Angiographic Substudy: The angiographic substudy is a parallel registry consisting of up to 30 additional consecutive subjects meeting all eligibility criteria treated with the SELUTION DEB recruited at select study sites. These subjects will undergo angiography at 12 months post-procedure. Clinical follow-up will not extend beyond 12 months in this cohort. Pharmacokinetic (pK) Substudy: The pKsubstudy is a parallel registry consisting of up to 20 additional consecutive subjects meeting all eligibility criteria treated with the SELUTION DEB recruited at select study sites. This study will be conducted under an approved substudy protocol and will include blood draws at regular intervals to characterize the pK plasma profile of sirolimus. Primary Endpoint: Target lesion failure (TLF), defined as the composite of cardiac death, target-vessel myocardial infarction (MI) or clinically-driven target lesion revascularization (TLR) at 12 months. MI includes spontaneous (Type 1) MI using the 4th Universal Definition of Myocardial Infarction (UDMI) and peri-procedural MI using the Society for Cardiac Angiography and Intervention (SCAI) definition.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
DOUBLE
Enrollment
960
After target lesion pre-dilatation and preparation , a SELUTION SLR 014 PTCA DEB study device should be selected with nominal diameter equal to the RVD of the target lesion (1:1 ratio) and length that allows approximately 2.00 mm longer than the proximal and distal edges of the target lesion (defined as the proximal and distal extent of predilation). If IVUS is performed to assess RVD, and DEB upsizing is deemed clinically necessary for optimal results, a maximum nominal DEB diameter of 3.0 mm is permitted. The SELUTION SLR 014 PTCA DEB should then be delivered, positioned in and deployed per instructions per use. The balloon should be inflated for 60 seconds provided that the patient can tolerate this duration. The minimum inflation time is 30 seconds.
For subjects randomized to the control group (DES), the treating physician will choose an FDA cleared DES and follow lesion preparation and stent deployment according to the Instructions per use (IFU) and institutional practices. The DES should be sized per IFU with respect to the vessel RVD. The use of IVUS or OCT is encouraged to guide optimal stent placement and assess final results. After DES deployment, additional balloon inflations should be performed as needed to obtain \< 30% residual diameter stenosis and resolve proximal or distal edge dissections greater than or equal to NHLBI grade B.
St. Bernards
Jonesboro, Arkansas, United States
CISD Sharp Chula Vista
Chula Vista, California, United States
Cedars Sinai
Los Angeles, California, United States
HCA Los Robles
Thousand Oaks, California, United States
Harbor UCLA
Torrance, California, United States
Colorado Heart and Vascular
Target lesion failure (TLF)
Target lesion failure (TLF) is defined as the composite of cardiac death, target-vessel myocardial infarction (MI) or clinically-driven target lesion revascularization (TLR) at 12 months. MI includes spontaneous (Type 1) MI using the 4th Universal Definition of Myocardial Infarction (UDMI) and peri-procedural MI using the Society for Cardiac Angiography and Intervention (SCAI) definition.
Time frame: 12 months
PK Sub-study Primary Endpoint 1
PK parameters of C(max)
Time frame: 6 months
PK Sub-study Primary Endpoint 2
PK parameters of T(max)
Time frame: 6 months
PK Sub-study Primary Endpoint 3
PK parameters of AUC(last)
Time frame: 6 months
PK Sub-study Primary Endpoint 4
PK parameters of MRT(last)
Time frame: 6 months
Secondary Endpoint 1
Composite of all-cause mortality, target vessel MI or clinically driven target lesion revascularization
Time frame: 12 months
Secondary Endpoint 2
Lesion success, defined as attainment of \< 30% residual stenosis of target lesion using any percutaneous method
Time frame: Up to 7 days
Secondary Endpoint 3
Procedure success, defined as attainment of \< 30% residual stenosis of the target lesion using the assigned study device only without the occurrence of in-hospital major adverse cardiac events (MACE), composite of all-cause death, MI or any clinically-driven TLR.
Time frame: Up to 7 days
Secondary Endpoint 4
Composite safety endpoint, defined as the patient-oriented composite of any death, any MI (spontaneous or peri-procedural), or any repeat revascularization.
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 5
All-Cause Mortality
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 6
Cardiovascular Mortality
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 7
MI (spontaneous MI using the 4th UDMI, peri-procedural MI using SCAI and Academic Research Consortium \[ARC\]-2 definitions)
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 8
Clinically-driven TLR
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 9
all TLR
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 10
Clinically-driven Target Vessel Revascularization (TVR)
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 11
all TVR
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 12
Non-target lesion revascularization
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 13
TLF
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 14
Target vessel failure (TVF), defined as a composite of: cardiac death, target vessel MI (spontaneous or peri-procedural) and any clinically-driven TVR
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 15
Stent or target lesion segment thrombosis (definite or probable) according to the ARC criteria for acute, subacute, late, very late and cumulative stent thrombosis
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
Secondary Endpoint 16
Bleeding Academic Research Consortium (BARC) class 2-5
Time frame: 12 months
Secondary Endpoint 17
Net adverse clinical events, defined as death, MI (spontaneous or peri-procedural), TVR, stent/target lesion segment thrombosis or bleeding (BARC types 2-5, assessed to 12 months)
Time frame: Up to 7 days and at 1 month, 6 months, 12 months, 2 years, 3 years, 4 years, and 5 years
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Golden, Colorado, United States
ClinRe
Thornton, Colorado, United States
Yale New Haven Hospital
New Haven, Connecticut, United States
MedStar Washington Hospital Center
Washington D.C., District of Columbia, United States
Manatee Memorial
Bradenton, Florida, United States
...and 60 more locations