The goal of this clinical trial is to compare measurements of geographic atrophy (GA) area between several types of imaging, in order to assess accuracy. The main question to answer is which imaging device provides measurements that are most similar to the standard of care device. Participants will be patients of a retina doctor at University Station Eye Clinic with geographic atrophy, and can expect to be in the study for 60-75 minutes.
This study is a single site pilot comparative imaging study in which enrolled participants will undergo dilated fundus exam as per standard of care. Study images will be obtained using Optos ultrawide field fundus autofluorescence (FAF) and both blue and green ultrawide field FAF with Zeiss Clarus. Autofluorescence (AF) imaging will also be performed on the Spectralis as it is currently the gold standard. The goal will be to obtain all images in 60-75 minutes over one visit. Images will be assessed, and area of geographic atrophy measured independently by two experienced readers.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
62
Spectralis is a scanning laser ophthalmoscope which uses a blue light excitation wavelength of 488nm and a 500nm barrier filter to produce FAF images. Spectralis images 20-55 degrees of the retina.
Optos is an ultrawide field imaging platform which images up to 200 degrees of the retina. It uses both a green-light excitation wavelength of 532nm and a red-light excitation wavelength of 633nm with an emission filter of greater than 540nm to produce FAF images.
Zeiss Clarus 700 is an ultrawide field imaging system with similar retinal coverage to that of Optos. It uses Broad Line Fundus Imaging to produce blue FAF images at excitation wavelengths of 435-500nm and green FAF images at wavelengths of 500-585nm
University of Wisconsin
Madison, Wisconsin, United States
Comparison of Geographic Atrophy Area Measurements With Heidelberg Spectralis to Zeiss Clarus 700 and Optos California
Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the devices.
Time frame: Up to 75 minutes
Comparison of Geographic Atrophy Area Measurements With Heidelberg Spectralis to Zeiss Clarus 700 and Optos California (Full Range Reported)
Graders will independently measure area of GA from images of each device. The measurements (in mm²) will be compared between the devices.
Time frame: Up to 75 minutes
Number of Eyes With Greater Than 10% Difference in Geographic Atrophy Area Than Heidelberg RegionFinder
Time frame: up to 75 mintues
Prevalence of Atrophy Outside the Standard 30 Degrees in Clarus and Optos Ultrawide Field Images
The presence of atrophy outside field 2 will be documented from the two ultrawide field imaging devices
Time frame: Up to 75 minutes
Prevalence of Other Autofluorescence Abnormalities Outside the Standard 30 Degrees in Clarus and Optos Ultrawide Field Images
The presence of abnormal autofluorescence outside field 2 will be documented from the two ultrawide field imaging devices
Time frame: Up to 75 minutes