This study is also designed to compare the performance of Masimo W1™ and Apple Watch Series 8 against FDA cleared hospital grade pulse oximetry technology under desaturation conditions.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
20
Noninvasive wearable health monitoring device
Human Pharmacology and Physiology Laboratory (HPPL) - Duke University Medical Center
Durham, North Carolina, United States
Accuracy of Masimo W1™ and Apple Watch Series 8 in Determining SpO2 (Oxygen Saturation) Values in Subjects During Desaturation
Accuracy will be determined by comparing Masimo W1™ SpO2 measurements and Apple Watch Series 8 SpO2 measurements with FDA-cleared pulse oximetry device SpO2 measurements and calculating the Arithmetic root mean square (ARMS) value. All values presented below for SpO2 (Bias, Precision and ARMS) are in units of %. The pairwise samples consist of paired SpO2 timing measurements from both the reference sensor, Masimo W1, and the Apple Watch.
Time frame: 1-6 hours
Successful Read Rate for Motion Conditions Under Slow Desaturation for Masimo W1™ and Apple Watch Series 8
This analysis examines the successful read rates for motion conditions during slow desaturation events for the Masimo W1™ and Apple Watch Series 8. The motion condition consists of a combined value representing three types of simulated motion: drinking, scratching the nose, and checking the time. Successful Read Rate = 1 - No Read Rate (Nf / Nspotcheck) x 100 (%) Nf = Number of Reads without Valid SpO2 Nspotcheck = Number of All Spot Checks
Time frame: 1-6 hours
Detection Rate of Fast Desaturation Events Under Various Conditions for Masimo W1™ and Apple Watch Series 8
Comparison of the detection rate of fast desaturation events under different conditions for the Masimo W1™ and Apple Watch Series 8. The first condition involves no motion, with the palm facing down and the Apple Watch in sleep mode, while the second condition involves no motion, with the palm facing sideways and manual spot checks performed on the Apple Watch. The goal is to evaluate the performance of both devices in detecting desaturation events under these varying circumstances. Detection Rate = (Nt / Ndesat) x 100 (%) Nt = Number of Detected Event by Test Device All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo W1™ and Apple Watch Series 8. Ndesat = Number of All Valid Fast Desaturation Events by Reference SpO2
Time frame: 1-6 hours
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.