The purpose of this study is to validate the accuracy of the CONNEQT Pulse blood pressure monitoring system for the measurement of blood pressure as compared to a gold standard, auscultatory sphygmomanometer in a pregnant population.
Clinical staff will review patient's medical record to determine eligibility. Following informed consent, a brief medical history will be collected. Serial blood pressure measurements will be collected to compare an auscultatory sphygmomanometer to blood pressure measurements of the CONNEQT Pulse blood pressure monitoring system in a pregnant population according to ISO protocol 81060-2.
Study Type
OBSERVATIONAL
Enrollment
41
Serial blood pressure measurements to compare reference auscultatory sphygmomanometer to CONNEQT Pulse according to ISO 81060-2.
Mount Carmel East
Columbus, Ohio, United States
Comparison of reference auscultatory blood pressure readings to CONNEQT Pulse blood pressure readings.
Paired means and standard deviations of both systolic and diastolic blood pressure readings will be analyzed according to section 5.2.4.1.2 of ISO 81060-2 protocol.
Time frame: Up to 60 minutes