Validation of hemodynamic measurements using the non-invasive Task Force Core(R) and Task Force Cardio(R) device with invasive hemodynamic measurements using the thermodilution technique (PICCO device).
The investigational device, Task Force® CORE (TFC, CNSystems Medizintechnik GmbH, Graz, Austria) together with the Task Force® CARDIO (TFCARDIO, CNSystems Medizintechnik GmbH, Graz, Austria) medical device software, is a new CE-marked medical device for continuous non-invasive determination of blood pressure (BP), cardiac output (CO) and derived parameters based on the well established CNAP® (continuous non-invasive arterial pressure) technology by CNSystems. The primary aim of this prospective, method comparison, open, multi-center study is data acquisition for the performance evaluation of the TFC in comparison with invasive reference methods (i.e. transpulmonary thermodilution, invasive blood pressure with radial artery catheter). The setting of the study will be the Operating Room (OR). This setting allows the comparison of the investigational device's cardiac output (CO) to invasive CO as well as of BP and derived hemodynamic parameters like Stroke Volume (SV), and Systemic Vascular Resistance (SVR). Measurements will be performed during the perioperative period, starting 15 minutes after the first skin incision.
Study Type
OBSERVATIONAL
Enrollment
70
Intraoperative cardiac output will be measured using the non-invasive device (Task Force CORE(R) and Task Force CARDIO(R) and will be compared with the invasive thermodilution method (PiCCO).
Medical University of Graz
Graz, Styria, Austria
Difference in intraoperative cardiac output measurements between Task Force CORE compared with thermodilution during surgery
Intraoperative performance of cardiac output values derived by the Task Force CORE system compared with cardiac output measurements derived by thermodilution measurements.
Time frame: For the duration of surgery
Differences in intraoperative hemodynamic measurements (blood pressure in mmHg) between Task Force CORE compared with thermodilution during surgery
Intraoperative evaluation of the performance of Task Force variable blood pressure (in mmHg) throughout surgery compared to simultaneous invasive measurements.
Time frame: For the duration of surgery
Differences in intraoperative hemodynamic measurements (stroke volume in ml) between Task Force CORE compared with thermodilution during surgery
Intraoperative evaluation of the performance of Task Force variable stroke volume (in ml) throughout surgery compared to simultaneous invasive measurements.
Time frame: For the duration of surgery
Differences in intraoperative hemodynamic measurements (stroke volume variation in %) between Task Force CORE compared with thermodilution during surgery
Intraoperative evaluation of the performance of Task Force variable stroke volume variation (in %) throughout surgery compared to simultaneous invasive measurements.
Time frame: For the duration of surgery
Intraoperative continuous Cardioc Output measurements
The comparison of intraoperative continuous cardiac output measurements derived by the Task Force device with continuous invasive cardiac output measurements derived by the PiCCO throughout surgery or before, during and after clinically indicated interventions (e.g. medication, passive leg raising, fluid challenge).
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: For the duration of surgery