The STYLE-LBBP study aims to compare the efficacy and safety of left-bundle branch pacing between the two types of available pacing leads: lumenless vs stylet-driven.
STYLE-LBBP is a randomized, non-blinded, multicenter clinical trial that compares the efficacy and safety of two types of available pacing leads in conduction system pacing (CSP). Patients referred to CSP who meet the inclusion and exclusion criteria will be enrolled and randomized into two groups: i)Left bundle branch pacing (LBBP) using leads without internal light (Lumenless; Medtronic Selectsecure 3830, Minneapolis, USA) or ii) LBBP with stylet-driven leads (Tendril STS 2088TC, Abbott, Inc., USA; Solia S60, Biotronik, SE \& Co., KG, Germany; Ingevity +, Boston Scientific, Marlborough, MA, USA). Efficacy in terms of achieving LBBP (selective and non-selective left bundle capture criteria) will be measured at implantation and after 24 hours. Complications of the procedure will also be recorded.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
120
Pacemaker implantation for conduction system pacing will be done using leads without internal lumen (Lumenless; Medtronic Selectsecure 3830, Minneapolis, USA).
Pacemaker implantation for conduction system pacing will be done using leads with internal lumen and retractable helix (Tendril STS 2088TC, Abbott, Inc., USA; Solia S60, Biotronik, SE \& Co., KG, Germany; Ingevity +, Boston Scientific, Marlborough, MA, USA)
Puerta de Hierro University Hospital
Majadahonda, Madrid, Spain
RECRUITINGFundación Jiménez Díaz University Hospital
Madrid, Spain
NOT_YET_RECRUITINGProportion of patients in which left-bundle branch pacing is achieved at implantation
Defined as a paced (pseudo) right bundle branch block QRS morphology with terminal r/R' in lead V1 and any of the following: i) recording of a LBB potential during intrinsic rhythm (only in patients with normal ventricular activation), ii) transition from non-selective to selective LBBP or non-selective LBBP to LV myocardial capture during decreasing pacing output, iii) abrupt shortening of stimulus to R wave peak time in V6 ECG lead (V6RWPT) ≥10 ms during implantation, iv) V6RWPT \< 75ms or \< 80ms in patients with preexisting left bundle branch block or v) an interpeak interval (V1RWPT - V6RWPT) ≥ 33ms.
Time frame: Baseline (During implantation)
Proportion of patients with left-bundle branch pacing criteria at 24 hours
Defined as a paced (pseudo) right bundle branch block QRS morphology with terminal r/R' in lead V1 and any of the following: i) recording of a LBB potential during intrinsic rhythm (only in patients with normal ventricular activation), ii) transition from non-selective to selective LBBP or non-selective LBBP to LV myocardial capture during decreasing pacing output, iii) abrupt shortening of stimulus to R wave peak time in V6 ECG lead (V6RWPT) ≥10 ms during implantation, iv) V6RWPT \< 75ms or \< 80ms in patients with preexisting left bundle branch block or v) an interpeak interval (V1RWPT - V6RWPT) ≥ 33ms.
Time frame: At 24 hours (before discharge)
Procedural complications
Number of patients with major and minor complications related to the procedure (death, perforation, cardiac tamponade, lead dislodgment, loss of capture or need for reintervention)
Time frame: At 24 hours