Assess the accuracy of INVSENSOR00069's peripheral oxygen saturation (SpO2) against contemporaneous measurement from arterial blood gas (ABG) analysis.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
OTHER
Masking
NONE
Enrollment
89
Noninvasive pulse oximeter
Masimo Corporation
Irvine, California, United States
SpO2 Accuracy of Masimo INVSENSOR00069
The SpO2 accuracy of Masimo INVSENSOR00069 was determined by calculating the accuracy root mean-squared (Arms) difference between the measured values (SpO2) to the reference values (SaO2) in accordance with ISO-80601-2-61. Arms= √(∑(i=1 to n) ((SpO2\_i-SaO2\_i )\^2 ))/n.