The purpose of this Post-Market Clinical Follow-up (PMCF) study is to describe the long-term safety and performance of the AcrySof IQ PanOptix IOL toric and non-toric models in subjects bilaterally implanted with these IOLs for 3 to 5 years. This study will be conducted in Spain.
Subjects will be recruited from a population that has already undergone lens implantation. Retrospective data will be collected from the pre-operative and surgical visits and any safety outcomes reported prior to enrollment. Prospective data will be collected at Visit 1, which will occur Year 3-5 postoperative.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
211
Trifocal toric IOL implanted in the capsular bag in the eye for the visual correction of aphakia and pre-existing corneal astigmatism in adult patients with and without presbyopia who desire near, intermediate and distance vision with increased spectacle independence.
Trifocal IOL implanted in the capsular bag in the eye for the visual correction of aphakia in adult patients with less than 1.00 diopter of preoperative corneal astigmatism with and without presbyopia who desire near, intermediate and distance vision with increased spectacle independence.
Hospital Universitari General de Catalunya
Sabadell, Barcelona, Spain
Oftalvist Jerez - HLA Puerta del Sur
Jerez de la Frontera, Cadiz, Spain
Barraquer Ophthalmology Center
Barcelona, Spain
Institut Catala de la Retina - Sarria
Barcelona, Spain
Mean Binocular Best Corrected Distance Visual Acuity (BCDVA)
BCDVA was assessed for both eyes together at a distance of 4 meters using letter charts and recorded in logarithm of the minimum angle of resolution (logMAR). The LogMAR VA scale ranges from -0.30 (20/10 Snellen) to 1.00 (20/200 Snellen). A LogMAR value of 0 equates to 20/20 Snellen VA (normal distance eyesight), with negative LogMAR values representing better than 20/20 VA. Subjects with at least one implanted eye with a toric IOL were categorized as toric as specified in the statistical analysis plan. No hypothesis testing was pre-specified for this endpoint.
Time frame: Year 3-5 postoperative (Visit 1, Day 1)
Number of Eyes Experiencing an Ocular Adverse Event by Toricity
Ocular adverse events were identified retrospectively from chart review of operative and immediate postoperative activities and prospectively at Visit 1 (Year 3-5 postoperative). Examples of ocular adverse events include the following: * Cystoid Macular Edema (CME) * Hypopyon * Endophthalmitis * Lens dislocation * Pupillary block * Retinal detachment * Secondary Surgical Interventions (explantation/exchange/repositioning) This outcome measure was pre-specified for eye-based reporting.
Time frame: Retrospective data from preoperative visit to Year 3-5 postoperative (Visit 1, Day 1)
Number of Subjects Experiencing a Non-Ocular Adverse Event
AEs were identified retrospectively from chart review of operative and immediate postoperative activities and prospectively at Visit 1 (Year 3-5 postoperative).
Time frame: Retrospective data from preoperative visit to Year 3-5 postoperative (Visit 1, Day 1)
Number of Device Deficiencies by Toricity
Device deficiencies were identified retrospectively from chart review of operative and immediate postoperative activities and prospectively at Visit 1 (Year 3-5 postoperative). Examples of device deficiencies include the following: * Failure to meet product specifications (e.g., incorrect IOL power) * IOL defect * Broken IOL optic * Broken IOL haptic * Scratched IOL optic * Unsealed device packaging * Suspected product contamination * Lack of performance This outcome measure was pre-specified for eye-based reporting.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Hospital Arruzafa
Córdoba, Spain
Miranza IOA
Madrid, Spain
Clinicas Novovision - Clinica Madrid
Madrid, Spain
Time frame: Retrospective data from preoperative visit to Year 3-5 postoperative (Visit 1, Day 1)