Rationale: Most BGM (blood glucose monitoring) devices in Vietnam are manufactured for home use only. However, they are commonly used in healthcare facility, primarily at point of care. We want to assess the validity of these devices by using CLSI (Clinical and Laboratory Standards Institute) POCT12-A3 to assess the accuracy. Objective: To validate the accuracy and effect of hematocrit on these devices, along with the precision profile using CLSI EP15A3 Study design: This is a cross-sectional study. Study population: Because we need a wide range of blood glucose level. We chose pregnant women who underwent 75g glucose tolerance test.
Most blood glucose monitors circulating the market are manufactured after ISO 15197:2013 standard, which is intended for self-measurement. Medical organizations need to verify blood glucose monitors if they are designed for prescription point-of-care use or validate blood glucose monitors if they are designed for over-the-counter use. Pregnant women are selected after consent, their blood are then drawn by phlebotomist into an EDTA (ethylenediamine tetraacetic acid) with NaF (sodium flouride) blood tube. Samples are run on alinity h hematology analyzer, manufactured by Abbott, to measure hematocrit. The analyzers are verified by laboratory procedure, recommended by CLSI (Clinical and Laboratory Standards Institute) method and quality control is checked before in accordance to the laboratory procedure. After that, blood tube is mixed thoroughly before being pipetted. Each BGM device will measure blood glucose level of the pipetted whole blood. Finally, samples will be run on a cobas c 502 analyzer to measure plasma glucose level, which is a reference method for determining the accuracy of the BGM devices. Statistical analysis: Meter performance is acceptable for use in hospitals when 95% of the individual results from the POC glucose meter system agree within 0.67 mmol/L (12 mg/dL) of the laboratory analyzer values at glucose concentrations below 5.55 mmol/L (100 mg/dL) and within 12.5% of the laboratory analyzer values at glucose concentrations at or above 5.55 mmol/L (100 mg/dL). In addition, the sum of the number of individual results with 1) errors that exceed 0.83 mmol/L (15 mg/dL) at glucose concentrations below 4.2 mmol/L (75 mg/dL) and 2) errors that exceed 20% at glucose concentrations at or above 4.2 mmol/L (75 mg/dL) should not exceed 2% of all results. An ordinary linear regression is performed to see how well the glucose level of BGM devices correlates with the reference method. Bland-Altman difference plots are used to describe the difference along with the hematocrit of each sample.
Study Type
OBSERVATIONAL
Enrollment
135
Hung Vuong Hospital
Ho Chi Minh City, Vietnam
Hematocrit
Hematocrit in percentage, measured on hematology analyzer alinity h
Time frame: Each participant's blood samples are run immediately after phlebotomy
Reference Glucose concentration
Glucose concentration, measured on reference glucose analyzer cobas c 502
Time frame: Each participant's blood samples are run immediately after phlebotomy
ACCU-CHEK Inform II glucose concentration
Glucose concentration, measured on assessed blood glucose monitoring device ACCU-CHEK Inform II, Serial number UU14463019
Time frame: Each participant's blood samples are run immediately after phlebotomy
ACCU-CHEK performa glucose concentration
Glucose concentration, measured on assessed blood glucose monitoring device ACCU-CHEK performa, Serial number 68827245972
Time frame: Each participant's blood samples are run immediately after phlebotomy
ONETOUCH VerioVue glucose concentration
Glucose concentration, measured on assessed blood glucose monitoring device ONETOUCH VerioVue, Serial number YEPMT06H
Time frame: Each participant's blood samples are run immediately after phlebotomy
Contour Plus glucose concentration
Glucose concentration, measured on assessed blood glucose monitoring device Contour Plus, Serial number 6128289
Time frame: Each participant's blood samples are run immediately after phlebotomy
EasyGlucose glucose concentration
Glucose concentration, measured on assessed blood glucose monitoring device EasyGlucose, Serial number G16D21I0700816
Time frame: Each participant's blood samples are run immediately after phlebotomy
VivaChek Ino glucose concentration
Glucose concentration, measured on assessed blood glucose monitoring device VivaChek Ino, Serial number 311Z0021DFA
Time frame: Each participant's blood samples are run immediately after phlebotomy
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.